Application of a g-C 3 N 4 @nZVI/CNTs Modified Carbon Paste Electrode for Uric Acid Detection in Spiked Urine Samples
ECS Sensors Plus · 01.12.2025 · 4 (4) · 043601
DOI: 10.1149/2754-2726/ae1bb7 ↗Kurz gefasst
Fachartikel in der Zeitschrift ECS Sensors Plus, online erschienen am 18.11.2025. Verfasst von Aya A. Mouhamed und 5 weiteren Autor:innen. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
A carbon paste electrode (CPE) based on a g-C 3 N 4 @nZVI/CNT nanocomposite was fabricated via chemical reduction and subsequently evaluated towards electrochemical detection of uric acid (UA) via differential pulse voltammetry (DPV). During synthesis, graphitic carbon nitride (g-C 3 N 4 ) serves as a durable substrate with extensive surface area promoting the homogeneous distribution of nano zero-valent iron (nZVI) particles. Carbon nanotubes (CNTs) were incorporated into the composite to enhance electrical conductivity and mechanical stability. The subsequent nanocomposite was characterized by scanning electron microscopy and energy-dispersive X-ray spectroscopy, confirming the successful integration of g-C 3 N 4 , nZVI, and CNTs with well-dispersed iron nanoparticles on the g-C 3 N 4 matrix. The resulting g-C 3 N 4 @nZVI/CNT-modified CPE demonstrated superior electrochemical sensitivity toward UA as evidenced by a pronounced increase in peak current during DPV analysis compared to bare and individually modified electrodes. Under optimized conditions, the sensor exhibited a linear response of 2.0–100.0 μM and detection limit of 1.7 μM. The electrode maintained high selectivity in the presence of common interferants with good repeatability and long-term operational stability. These findings highlight the synergistic interaction among g-C 3 N 4 , nZVI, and CNTs in boosting electron transfer and catalytic activity, making this composite a promising, cost-effective, and scalable solution for sensitive UA detection in clinical diagnostics and environmental monitoring.
Zitieren
APA
Mouhamed, A. A., Hussein, O. G., Mekhail, M. O., Zaki, T. V., Mahmoud, A. M., & Bell, J. G. (2025). Application of a g-C 3 N 4 @nZVI/CNTs Modified Carbon Paste Electrode for Uric Acid Detection in Spiked Urine Samples. ECS Sensors Plus, 4(4), 043601. https://doi.org/10.1149/2754-2726/ae1bb7
Vancouver
Mouhamed AA, Hussein OG, Mekhail MO, Zaki TV, Mahmoud AM, Bell JG. Application of a g-C 3 N 4 @nZVI/CNTs Modified Carbon Paste Electrode for Uric Acid Detection in Spiked Urine Samples. ECS Sensors Plus. 2025;4(4):043601. doi:10.1149/2754-2726/ae1bb7
BibTeX
@article{Mouhamed2025,
author = {Mouhamed, Aya A. and Hussein, Ola G. and Mekhail, Maria Osama and Zaki, Tony Victor and Mahmoud, Amr M. and Bell, Jeffrey G.},
title = {Application of a g-C 3 N 4 @nZVI/CNTs Modified Carbon Paste Electrode for Uric Acid Detection in Spiked Urine Samples},
journal = {ECS Sensors Plus},
year = {2025},
volume = {4},
number = {4},
pages = {043601},
doi = {10.1149/2754-2726/ae1bb7}
}