Usability assessment of Versius, a new robot-assisted surgical device for use in minimal access surgery
BMJ Surgery, Interventions, & Health Technologies · 2020-05 · 2 (1) · e000028
DOI: 10.1136/bmjsit-2019-000028 ↗Kurz gefasst
Fachartikel in der Zeitschrift BMJ Surgery, Interventions, & Health Technologies, online erschienen am 24.05.2020. Verfasst von Fiona Haig und 3 weiteren Autor:innen. Beteiligte Einrichtung: CMR Surgical, Cambridge, Cambridgeshire, UK. Die Arbeit wurde bisher 67-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: CMR Surgical.
Abstract (Originalsprache)
Objectives
Versius is a teleoperated surgical robotic system intended for use in minimal access surgery. This study aimed to validate the safety and effectiveness of the Versius user interface in the hands of trained users and identify and address the causes of any user errors.
Design
Surgical teams completed a commercially representative training program over 3.5 days. After training was completed, the usability-related aspects of the system were assessed.
Setting
A simulated operating room using a cadaveric model.
Participants
Surgical teams consisting of a lead surgeon, assistant surgeon, scrub nurse and circulating nurse. Main outcome measures Usability-related aspects of the system were assessed through the completion of critical and non-critical tasks. A critical task was defined according to the Food and Drug Administration’s definition, as a user task which, if performed incorrectly or not performed at all, would or could cause serious harm to the patient or user, where harm is defined to include compromised medical care.
Results
In total, 17 surgical teams participated in the study and all were experienced in laparoscopic surgery. The number of robotic surgeries performed by the participants per month ranged from 0 to 100. Surgical specialties were similarly represented from obstetrics and gynecology, colorectal, urology and upper gastrointestinal. No critical task failures were observed. Of all the tasks completed, 98% were recorded as a pass or a pass with difficulty.
Conclusions
These results demonstrate that in a simulated clinical setting, Versius can be safely used by both laparoscopically and robotically trained healthcare professionals. These results support the progression to assessment of Versius in preclinical studies.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- CMR Surgical, Cambridge, Cambridgeshire, UK.
ORCID-Profile: Mark Slack
Förderung
- CMR Surgical
Zitieren
APA
Haig, F., Medeiros, A. C. B., Chitty, K., & Slack, M. (2020). Usability assessment of Versius, a new robot-assisted surgical device for use in minimal access surgery. BMJ Surgery, Interventions, & Health Technologies, 2(1), e000028. https://doi.org/10.1136/bmjsit-2019-000028
Vancouver
Haig F, Medeiros ACB, Chitty K, Slack M. Usability assessment of Versius, a new robot-assisted surgical device for use in minimal access surgery. BMJ Surgery, Interventions, & Health Technologies. 2020;2(1):e000028. doi:10.1136/bmjsit-2019-000028
BibTeX
@article{Haig2020,
author = {Haig, Fiona and Medeiros, Ana Cristina Barbosa and Chitty, Karen and Slack, Mark},
title = {Usability assessment of Versius, a new robot-assisted surgical device for use in minimal access surgery},
journal = {BMJ Surgery, Interventions, & Health Technologies},
year = {2020},
volume = {2},
number = {1},
pages = {e000028},
doi = {10.1136/bmjsit-2019-000028}
}
EUVIMED