Physics-Informed In-Silico Dynamic Computed Tomography of Human Lungs: Generation, Evaluation, and Refinement

Sunder Neelakantan, Kyle J Myers, Reza Avazmohammadi iD

Journal of Biomechanical Engineering · 01.10.2025 · 147 (10)

DOI: 10.1115/1.4069391 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Biomechanical Engineering, online erschienen am 29.08.2025. Verfasst von Sunder Neelakantan und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Texas A&M University, College Station, TX 77843. Hauptthemen laut MEDLINE: Lung, Four-Dimensional Computed Tomography, Computer Simulation, Tomography, X-Ray Computed. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NHLBI NIH HHS, National Science Foundation, National Heart, Lung, and Blood Institute.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Lung injuries lead to heterogeneous ventilation behavior in the lung parenchyma, and conventional methods used to assess lung health, such as spirometry, fall short of providing regional information about lung function. Dynamic medical imaging and image registration offer a powerful tool for estimating the kinematic behavior of lung parenchyma in vivo. However, the difficulty of validating lung deformation estimated by image registration has curbed widespread adoption in the clinic. In-silico images, reconstructed from finite element (FE) simulations, provide a method to verify the results estimated through image registration (IR). Our objective in this study was to use in-silico computed tomography (CT) images, reconstructed from FE simulations, to assess the accuracy of an image registration method. In this study, we used dynamic CT (4DCT) images from human patients to reconstruct the lungs and generate an FE mesh. In-silico simulations were performed using the lung FE mesh, and the results were used to generate in-silico dynamic CT images matching the resolution of the actual 4DCT images. Image registration was performed on the actual and in-silico images, and the results were compared to those from the FE simulation. Results indicated good agreement in displacement estimated by the FE simulations and the image registration of the actual and in-silico CT images. The difference in predicted displacement image registration of the actual CT images and the FE simulations was greatest at the main bronchi, with a value of 2.7 mm. This result highlighted the effectiveness of the FE simulation-based method to generate in-silico CT images. The volumetric strain comparisons between actual 4DCT and the in-silico images were used to assess the method's accuracy. A new set of in-silico images was generated at a higher spatial resolution, resulting in improved agreement for the volumetric strain contours. We expect the method reported in this study to be applied to optimize medical imaging methods and investigate the behavior of various lung diseases under medical imaging.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LungFour-Dimensional Computed TomographyComputer SimulationTomography, X-Ray ComputedHumansFinite Element Analysis

Stichwörter der Autor:innen

computed tomographyimage registrationin-silico imageslungsphysics-informed motion registration

Autor:innen und Institutionen

  • Department of Biomedical Engineering, Texas A&M University, College Station, TX 77843.
  • Texas A&M University.
  • Hagler Institute for Advanced Study, Texas A&M University, College Station, TX 77840.
  • Department of Biomedical Engineering, Texas A&M University, College Station, TX 77843; J. Mike Walker '66 Department of Mechanical Engineering, Texas A&M University, College Station, TX 77840; Department of Cardiovascul…

ORCID-Profile: Reza Avazmohammadi

Förderung

  • NHLBI NIH HHS
  • National Science Foundation
  • National Heart, Lung, and Blood Institute

Zitieren

APA

Neelakantan, S., Myers, K. J., & Avazmohammadi, R. (2025). Physics-Informed In-Silico Dynamic Computed Tomography of Human Lungs: Generation, Evaluation, and Refinement. Journal of Biomechanical Engineering, 147(10). https://doi.org/10.1115/1.4069391

Vancouver

Neelakantan S, Myers KJ, Avazmohammadi R. Physics-Informed In-Silico Dynamic Computed Tomography of Human Lungs: Generation, Evaluation, and Refinement. Journal of Biomechanical Engineering. 2025;147(10). doi:10.1115/1.4069391

BibTeX
@article{Neelakantan2025,
  author = {Neelakantan, Sunder and Myers, Kyle J and Avazmohammadi, Reza},
  title = {Physics-Informed In-Silico Dynamic Computed Tomography of Human Lungs: Generation, Evaluation, and Refinement},
  journal = {Journal of Biomechanical Engineering},
  year = {2025},
  volume = {147},
  number = {10},
  doi = {10.1115/1.4069391}
}
Kostenlos starten →