EUVIMEDEUVIMED entdecken
Crossref

Design and Development of Patient-Specific Miniplate for the Treatment of Maxillofacial Fractures—A Finite Element Study

Sandeep Choudhury, Anik Banerjee, Santanu Majumder, Amit Roy Chowdhury

Journal of Engineering and Science in Medical Diagnostics and Therapy · 01.08.2023 · 6 (3)

DOI: 10.1115/1.4062509 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Engineering and Science in Medical Diagnostics and Therapy, online erschienen am 07.06.2023. Verfasst von Sandeep Choudhury und 3 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 65 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Fractures of the maxillofacial region are prevalent in both developed and developing nations. Maxillofacial fractures primarily occur as a result of incidents such as car crashes, physical assaults, and attacks. Although using miniplates to treat maxillofacial fractures is a widely accepted practice, the challenge lies in selecting the appropriate miniplate size that is tailored to the specific dimensions of the fracture in each patient. The study aims to evaluate and determine the most suitable design between a two-hole miniplate and a four-hole miniplate for securing a fractured mandible, with the ultimate goal of creating a customized solution for each patient. A mandible model is chosen with a diagonal length of 82 mm and created perpendicular fracture gaps measuring 1 mm to 3 mm on the left buccolingual section of the solid model. A two-hole and a four-hole miniplate are designed with different thicknesses (ranging from 1 mm to 2 mm) and gap distances (ranging from 6 mm to 12 mm). The miniplates were put together in a model of the mandible. To test their effectiveness, the combination of the mandible model and plates was subjected to various muscle forces, as well as the force of biting, using finite element analysis. At the mandible location, the four-hole miniplate implantation exhibits superior stabilization in comparison to the two-hole miniplate assembly. The miniplate's size and dimensions can be adjusted depending on the size of the fracture in the mandible, resulting in a patient-specific solution for the implantation of miniplate in the mandible.

Zitieren

APA

Choudhury, S., Banerjee, A., Majumder, S., & Chowdhury, A. R. (2023). Design and Development of Patient-Specific Miniplate for the Treatment of Maxillofacial Fractures—A Finite Element Study. Journal of Engineering and Science in Medical Diagnostics and Therapy, 6(3). https://doi.org/10.1115/1.4062509

Vancouver

Choudhury S, Banerjee A, Majumder S, Chowdhury AR. Design and Development of Patient-Specific Miniplate for the Treatment of Maxillofacial Fractures—A Finite Element Study. Journal of Engineering and Science in Medical Diagnostics and Therapy. 2023;6(3). doi:10.1115/1.4062509

BibTeX
@article{Choudhury2023,
  author = {Choudhury, Sandeep and Banerjee, Anik and Majumder, Santanu and Chowdhury, Amit Roy},
  title = {Design and Development of Patient-Specific Miniplate for the Treatment of Maxillofacial Fractures—A Finite Element Study},
  journal = {Journal of Engineering and Science in Medical Diagnostics and Therapy},
  year = {2023},
  volume = {6},
  number = {3},
  doi = {10.1115/1.4062509}
}
Kostenlos starten →