Preventing overfilling of blood bags to reduce clot formation and improve donor safety
Vox Sanguinis · 2026-08 · 121 (8) · 1256-1261
DOI: 10.1111/vox.70293 ↗At a glance
Journal article in Vox Sanguinis, published online on 2026-05-21. Written by Kabiru M Sulaiman and 6 co-authors. Institution: Department of Haematology and Blood Transfusion, Aminu Kano Teaching Hospital, Kano, Nigeria. Major MEDLINE topics: Blood Donors, Blood Safety, Blood Banks, Thrombosis. 14 references. The full text is freely available (open access, CC BY 4.0). Funded by: Deutsche Gesellschaft für Internationale Zusammenarbeit, Else Kröner-Fresenius-Stiftung, German Federal Ministry for Economic Cooperation and Development (BMZ).
Abstract (original language)
Background and Objectives Overfilling of blood bags is common in low‐resource settings, causing clot formation, donor and recipient risks and blood wastage. This study assessed overfilling prevalence and evaluated a low‐cost intervention to mitigate it. Materials and Methods We conducted a prospective study across six blood banks in Kano, Nigeria, weighing 131 blood bags and examining clot formation in 688 consecutive units. A simple, manual measuring and mixing (MeMi) device was developed and implemented to prevent overfilling.
Results
Before intervention, 129/131 blood bags (98.5%) exceeded the 612 g upper limit (mean: 711 g; range: 560–888 g). Systematic assessment revealed macroclots in 72/688 bags (10.5%; 95% confidence interval [CI]: 8.18–12.76). After MeMi use, 658/723 bags (91%) met the target weight (500–612 g), while 2 were overweight (629 g, 639 g) and 63 underweight (265–499 g). Clot incidence dropped to 1.9% (95% CI: 0%–5.1%).
Conclusion
Overfilling of blood bags poses serious safety concerns in low‐resource settings. The MeMi device provides a practical, low‐cost mechanical solution without needing electricity. Locally producible for ~€15, it significantly reduces overfilling and clotting. This improves safety for donors and recipients. Its simplicity and affordability make it highly scalable across regions like sub‐Saharan Africa.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Haematology and Blood Transfusion, Aminu Kano Teaching Hospital, Kano, Nigeria.
- National Blood Service Agency, Abuja, Nigeria.
- Institut für Transfusionsmedizin, Universitätsmedizin Greifswald, Greifswald, Germany.
- Medical School of Bayero University, Kano, Nigeria.
ORCID profiles: Andreas Greinacher
Funding
- Deutsche Gesellschaft für Internationale Zusammenarbeit
- Else Kröner-Fresenius-Stiftung
- German Federal Ministry for Economic Cooperation and Development (BMZ)
- Bundesministerium für Wirtschaftliche Zusammenarbeit und Entwicklung
Cite
APA
Sulaiman, K. M., Ismail, H., Ibrahim, Y. Y., Selleng, K., Gwarzo, A. K., Gwarzo, D., & Greinacher, A. (2026). Preventing overfilling of blood bags to reduce clot formation and improve donor safety. Vox Sanguinis, 121(8), 1256-1261. https://doi.org/10.1111/vox.70293
Vancouver
Sulaiman KM, Ismail H, Ibrahim YY, Selleng K, Gwarzo AK, Gwarzo D, et al. Preventing overfilling of blood bags to reduce clot formation and improve donor safety. Vox Sanguinis. 2026;121(8):1256-1261. doi:10.1111/vox.70293
BibTeX
@article{Sulaiman2026,
author = {Sulaiman, Kabiru M and Ismail, Halima and Ibrahim, Yahaya Yazid and Selleng, Kathleen and Gwarzo, Aisha K and Gwarzo, Dalhat and Greinacher, Andreas},
title = {Preventing overfilling of blood bags to reduce clot formation and improve donor safety},
journal = {Vox Sanguinis},
year = {2026},
volume = {121},
number = {8},
pages = {1256--1261},
doi = {10.1111/vox.70293}
}