Fachartikel in der Zeitschrift Vox Sanguinis, 2025. Verfasst von Sarah Goatson und 8 weiteren Autor:innen. Beteiligte Einrichtung: Welsh Blood Service, Pontyclun, UK. Hauptthemen laut MEDLINE: Blood Platelets, Blood Preservation, Platelet Transfusion. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 36 Literaturangaben. Gefördert von: University of the West of England, Welsh Blood Service.
Abstract (Originalsprache)
Background and Objectives Pre‐hospital critical care faces challenges in carrying platelet concentrates (PC). Cold‐stored platelets (CSP) may offer improved haemostatic function over room temperature platelets. Rapid warming of blood products is crucial for hypothermic trauma patients. This study evaluates whether the°M Warmer System (°MEQU, Copenhagen), used for red cells and intravenous thrombin receptor activator peptide 6 fluids, affects CSP functionality compared to a standard giving set. Materials and Methods Buffy coat‐derived platelets, in a platelet additive solution to plasma ratio of approximately 65:35, were stored at 4 ± 2°C without agitation for 6 days. A baseline measurement was taken on Day 2. On Day 6, a ‘pre‐sample’ was taken; then CSP were either infused through the°MEQU or a standard transfusion‐giving set. Platelet function was assessed by aggregation and thromboelastography (TEG). Platelet activation was assessed by CD62P expression, Annexin V binding, microparticle concentration and soluble CD62P concentration. Ten replicates were completed.
Results
Neither aggregation nor TEG showed significant differences between PC infused through the standard set or the rapid blood warmer. Soluble CD62P levels were significantly higher in both the standard giving set (26.7 ± 6.4 ng/mL) and rapid blood warmer (25.9 ± 5.1 ng/mL) compared to the pre‐sample (18.8 ± 4.8 ng/mL), with no significant difference between the groups.
Conclusion
The°MEQU did not detrimentally affect platelet quality and functionality compared to a standard giving set, suggesting the feasibility for a CSP clinical trial in the pre‐hospital environment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Welsh Blood Service, Pontyclun, UK.
- Emergency Medical Retrieval and Transfer Service (EMRTS) Cymru, Wales, UK.
- Centre of Defence Pathology, Royal Centre for Defence Medicine, Birmingham, UK.
ORCID-Profile: Jamie Nash, Christine Saunders, Edward J Sayers, David Rawlinson
Förderung
- University of the West of England
- Welsh Blood Service
Zitieren
APA
Goatson, S., Nash, J., Saunders, C., Pearce, N., Sayers, E. J., Rawlinson, D., Hingston, C., Scorer, T., & George, C. E. (2025). Impact of a rapid blood warmer on the quality and function of cold‐stored platelets. Vox Sanguinis. https://doi.org/10.1111/vox.70094
Vancouver
Goatson S, Nash J, Saunders C, Pearce N, Sayers EJ, Rawlinson D, et al. Impact of a rapid blood warmer on the quality and function of cold‐stored platelets. Vox Sanguinis. 2025. doi:10.1111/vox.70094
BibTeX
@article{Goatson2025,
author = {Goatson, Sarah and Nash, Jamie and Saunders, Christine and Pearce, Nicola and Sayers, Edward J and Rawlinson, David and Hingston, Christopher and Scorer, Thomas and George, Chloe E},
title = {Impact of a rapid blood warmer on the quality and function of cold‐stored platelets},
journal = {Vox Sanguinis},
year = {2025},
doi = {10.1111/vox.70094}
}
EUVIMED