Evaluation of platelets componentized with the Reveos Automated Blood Processing System and stored for 7 days at room temperature in a non‐Di(2‐ethylhexyl) phthalate (DEHP) platelet pooling set
Transfusion · 2024
DOI: 10.1111/trf.17780 ↗Kurz gefasst
Fachartikel in der Zeitschrift Transfusion, 2024. Verfasst von Kristin M Reddoch-Cardenas und 3 weiteren Autor:innen. Beteiligte Einrichtung: Blood and Shock Resuscitation, U.S. Army Institute of Surgical Research, JBSA-FT, Sam Houston, Texas, USA. Hauptthemen laut MEDLINE: Blood Preservation, Blood Platelets, Temperature. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: Combat Casualty Care Research Program.
Abstract (Originalsprache)
Background
Platelet concentrates (PCs) used for transfusion can be produced by apheresis or derived from whole blood (WB). The Reveos device is the first US Food and Drug Administration‐approved automated blood processing system that can produce PCs. In this work, we evaluated the quality and function of Reveos‐collected PCs stored for 7 days at room temperature. Study Design and Methods WB was collected from healthy donors and componentized on the day of collection (Fresh) or after an overnight hold (Overnight). PCs were produced ( n = 7 Fresh; n = 6 Overnight), stored at room temperature in plasma, and evaluated on days 1 and 7 for quality metrics, platelet activation, clot formation, and aggregation response.
Results
Platelet count was comparable between Fresh and Overnight PCs. A drop in pH was reported in Fresh day 7 PCs ( p < .001, vs. day 1) but not in Overnight. Overnight units displayed the lowest levels of P‐selectin expression ( p = .0008, vs. day 7 Fresh). Reduced clot strength and increased lysis were observed in both Fresh and Overnight units on day 7 (vs. day 1). Overnight‐hold PCs resulted in the highest clot strength on day 7 ( p = .0084, vs. Fresh). No differences in aggregation were reported between groups.
Conclusion
Reveos‐processed PCs produced from overnight‐hold WB performed better in hemostatic function assays and displayed reduced activation compared to fresh WB‐derived PCs, although both PC groups maintained platelet quality throughout storage. Utilization of overnight WB for PC preparation with Reveos holds promise as an alternative method of producing platelets for transfusion purposes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Blood and Shock Resuscitation, U.S. Army Institute of Surgical Research, JBSA-FT, Sam Houston, Texas, USA.
ORCID-Profile: Kristin M Reddoch-Cardenas
Förderung
- Combat Casualty Care Research Program
Zitieren
APA
Reddoch-Cardenas, K. M., Perez, S. L., Ferdin, J. N., & Meledeo, M. A. (2024). Evaluation of platelets componentized with the Reveos Automated Blood Processing System and stored for 7 days at room temperature in a non‐Di(2‐ethylhexyl) phthalate (DEHP) platelet pooling set. Transfusion. https://doi.org/10.1111/trf.17780
Vancouver
Reddoch-Cardenas KM, Perez SL, Ferdin JN, Meledeo MA. Evaluation of platelets componentized with the Reveos Automated Blood Processing System and stored for 7 days at room temperature in a non‐Di(2‐ethylhexyl) phthalate (DEHP) platelet pooling set. Transfusion. 2024. doi:10.1111/trf.17780
BibTeX
@article{ReddochCardenas2024,
author = {Reddoch-Cardenas, Kristin M and Perez, Samantha L and Ferdin, Justin N and Meledeo, Michael A},
title = {Evaluation of platelets componentized with the Reveos Automated Blood Processing System and stored for 7 days at room temperature in a non‐Di(2‐ethylhexyl) phthalate (DEHP) platelet pooling set},
journal = {Transfusion},
year = {2024},
doi = {10.1111/trf.17780}
}
EUVIMED