Intra‐Rater Variability and Inter‐Rater Reliability in Measuring Tear Osmolarity in Dogs Using the ScoutPro System

Joschka Spornberger iD, Petr Soukup iD, Andrea Meyer-Lindenberg iD, Ingrid Allgoewer iD

Veterinary Ophthalmology · 2025

DOI: 10.1111/vop.70053 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Veterinary Ophthalmology, 2025. Verfasst von Joschka Spornberger und 3 weiteren Autor:innen. Beteiligte Einrichtung: Animal Eye Practice, Berlin, Germany. Hauptthemen laut MEDLINE: Tears, Diagnostic Techniques, Ophthalmological. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 28 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To evaluate the applicability of the ScoutPro (Trukera Medical, Southlake, Texas, USA) osmolarity system for tear film osmolarity measurement in healthy dogs. Animals Studied Seventeen dogs (34 eyes) of dolichocephalic and mesocephalic breeds were tested. Procedures Tear osmolarity was tested with the ScoutPro osmolarity system. Two examiners performed three subsequent measurements of each eye. Intra‐rater variability and inter‐rater reliability were calculated.

Results

The technique was straightforward; instrument handling was simple. Mean tear osmolarity measured by examiner one was 323.51 ± 9.89 mOsm/L and by examiner two 326.74 ± 9.02 mOsm/L. Mean difference between examiners was 3.2 ± 11.2 mOsm/L. Paired two‐tailed t ‐test was p = 0.1. The intra‐rater variability was acceptable with a Cronbach's Alpha of 0.75 for rater one and 0.80 for rater two. There were no significant differences in the three measurements of the individual examiners. There were no significant differences in the three measurements of the individual examiners (asymptotic significance rater one was 0.56, asymptotic significance rater two was 0.11).

Conclusions

Measurements with the ScoutPro osmolarity system in dogs are easy to perform. Variability and inter‐rater reliability seem to be acceptable in the small group of dogs tested. Further studies regarding the reliability of the test results and the evaluation of the tear osmolarity in ocular surface disease are warranted.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TearsDiagnostic Techniques, OphthalmologicalAnimalsDogsOsmolar ConcentrationObserver VariationReproducibility of ResultsFemaleMale

Stichwörter der Autor:innen

I‐PEN VetTearLabcaninedry eye diseaseocular surfacetear film

Autor:innen und Institutionen

  • Animal Eye Practice, Berlin, Germany.
  • LMU Munich, Clinic for Small Animal Surgery and Reproduction, Munich, Germany.

ORCID-Profile: Joschka Spornberger, Petr Soukup, Andrea Meyer-Lindenberg, Ingrid Allgoewer

Zitieren

APA

Spornberger, J., Soukup, P., Meyer-Lindenberg, A., & Allgoewer, I. (2025). Intra‐Rater Variability and Inter‐Rater Reliability in Measuring Tear Osmolarity in Dogs Using the ScoutPro System. Veterinary Ophthalmology. https://doi.org/10.1111/vop.70053

Vancouver

Spornberger J, Soukup P, Meyer-Lindenberg A, Allgoewer I. Intra‐Rater Variability and Inter‐Rater Reliability in Measuring Tear Osmolarity in Dogs Using the ScoutPro System. Veterinary Ophthalmology. 2025. doi:10.1111/vop.70053

BibTeX
@article{Spornberger2025,
  author = {Spornberger, Joschka and Soukup, Petr and Meyer-Lindenberg, Andrea and Allgoewer, Ingrid},
  title = {Intra‐Rater Variability and Inter‐Rater Reliability in Measuring Tear Osmolarity in Dogs Using the ScoutPro System},
  journal = {Veterinary Ophthalmology},
  year = {2025},
  doi = {10.1111/vop.70053}
}
Kostenlos starten →