Analytical performance of feline plasma on current Heska and IDEXX point‐of‐care biochemistry analyzers compared with a commercial laboratory analyzer
Veterinary Clinical Pathology · 2024
DOI: 10.1111/vcp.13310 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Veterinary Clinical Pathology, 2024. Verfasst von Randolph M Baral und 3 weiteren Autor:innen. Beteiligte Einrichtung: University of Sydney/Paddington Cat Hospital, Sydney, Australia. Hauptthemen laut MEDLINE: Point-of-Care Systems, Blood Chemical Analysis. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
Point‐of‐care (POC) biochemistry analyzers are widely used in small animal clinical practice but infrequently independently assessed for performance.
Objective
To assess the performance of two current model point‐of‐care biochemistry analyzers (Heska Element DC and IDEXX Catalyst) compared with a commercial laboratory analyzer (Cobas 8000).
Methods
One hundred twenty‐one cats from a feline hospital population were sampled with plasma results from a single lithium heparin tube assessed on all three analyzers. Plasma biochemistry results from each POC analyzer were compared with the commercial laboratory analyzer using Bland–Altman difference plots and by determining whether the limits of agreement (LOAs) (95% of differences) fell within various quality goals after correcting for inherent bias.
Results
Only 7 of 14 analytes on the Heska analyzer and 2 analytes on the IDEXX analyzer attained the most stringent LOA quality goal, which was being within desirable total error based on biologic variation (TE des ). The number of analytes achieving quality goals increased with less stringent standards such as American Society of Veterinary Clinical Pathologists allowable total error (ASVCP TE A ) guidelines or if
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University of Sydney/Paddington Cat Hospital, Sydney, Australia.
- Vetnostics, Sydney, Australia.
- Veterinary Information Network, Davis, California, USA.
ORCID-Profile: Randolph M Baral, Susan M Jaensch, Douglas A Hayward, Kathleen P Freeman
Zitieren
APA
Baral, R. M., Jaensch, S. M., Hayward, D. A., & Freeman, K. P. (2024). Analytical performance of feline plasma on current Heska and IDEXX point‐of‐care biochemistry analyzers compared with a commercial laboratory analyzer. Veterinary Clinical Pathology. https://doi.org/10.1111/vcp.13310
Vancouver
Baral RM, Jaensch SM, Hayward DA, Freeman KP. Analytical performance of feline plasma on current Heska and IDEXX point‐of‐care biochemistry analyzers compared with a commercial laboratory analyzer. Veterinary Clinical Pathology. 2024. doi:10.1111/vcp.13310
BibTeX
@article{Baral2024,
author = {Baral, Randolph M and Jaensch, Susan M and Hayward, Douglas A and Freeman, Kathleen P},
title = {Analytical performance of feline plasma on current Heska and IDEXX point‐of‐care biochemistry analyzers compared with a commercial laboratory analyzer},
journal = {Veterinary Clinical Pathology},
year = {2024},
doi = {10.1111/vcp.13310}
}
EUVIMED