In vitro evaluation of a new viscoelastometry‐based point‐of‐care analyzer
Transfusion · 2024
DOI: 10.1111/trf.17808 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Transfusion, 2024. Verfasst von Vanina Corominas und 5 weiteren Autor:innen. Beteiligte Einrichtung: French Military Blood Institute, Clamart, France. Hauptthemen laut MEDLINE: Point-of-Care Systems. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 38 Literaturangaben.
Abstract (Originalsprache)
Introduction
The VCM is a point‐of‐care analyzer using a new viscoelastometry technique for rapid assessment of hemostasis on fresh whole blood. Its characteristics would make it suitable for use in austere environments. The purpose of this study was to evaluate the VCM in terms of repeatability, reproducibility and interanalyzer correlation, reference values in our population, correlation with standard coagulation assays and platelet count, correlation with the TEG5000 analyzer and resistance to stress conditions mimicking an austere environment.
Methods
Repeatability, reproducibility, and interanalyzer correlation were performed on quality control samples ( n = 10). Reference values were determined from blood donor samples ( n = 60). Correlations with standard biological assays were assessed from ICU patients ( n = 30) and blood donors ( n = 60) samples. Correlation with the TEG5000 was assessed from blood donor samples. Evaluation of vibration resistance was performed on blood donor ( n = 5) and quality control ( n = 5) samples.
Results
The CVs for repeatability and reproducibility ranged from 0% to 11%. Interanalyzer correlation found correlation coefficients ( r 2 ) ranging from 0.927 to 0.997. Our reference values were consistent with those provided by the manufacturer. No robust correlation was found with conventional coagulation tests. The correlation with the TEG5000 was excellent with r 2 ranging from 0.75 to 0.92. Resistance to stress conditions was excellent.
Conclusion
The VCM analyzer is a reliable, easy‐to‐use instrument that correlates well with the TEG5000. Despite some logistical constraints, the results suggest that it can be used in austere environments. Further studies are required before its implementation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- French Military Blood Institute, Clamart, France.
- French Military Health Service Academy, Ecole du Val-de-Grâce, Paris, France.
- Department of Anesthesiology and Intensive Care, Percy Military Medical Center, Clamart, France.
- Department of Laboratory Medicine, Percy Military Medical Center, Clamart, France.
ORCID-Profile: Vanina Corominas, Christophe Martinaud
Zitieren
APA
Corominas, V., Chiniard, T., Pasquier, P., Foissaud, V., de Rudnicki, S., & Martinaud, C. (2024). In vitro evaluation of a new viscoelastometry‐based point‐of‐care analyzer. Transfusion. https://doi.org/10.1111/trf.17808
Vancouver
Corominas V, Chiniard T, Pasquier P, Foissaud V, de Rudnicki S, Martinaud C. In vitro evaluation of a new viscoelastometry‐based point‐of‐care analyzer. Transfusion. 2024. doi:10.1111/trf.17808
BibTeX
@article{Corominas2024,
author = {Corominas, Vanina and Chiniard, Thomas and Pasquier, Pierre and Foissaud, Vincent and de Rudnicki, Stéphane and Martinaud, Christophe},
title = {In vitro evaluation of a new viscoelastometry‐based point‐of‐care analyzer},
journal = {Transfusion},
year = {2024},
doi = {10.1111/trf.17808}
}
EUVIMED