A prospective, pragmatic non‐inferiority study of emergency intubation success with the single‐use i‐view versus standard reusable video laryngoscope
Transfusion · 2024
DOI: 10.1111/trf.17790 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Transfusion, 2024. Verfasst von Steven G Schauer und 10 weiteren Autor:innen. Beteiligte Einrichtung: US Army Institute of Surgical Research, JBSA Fort Sam Houston, Texas, USA. Hauptthemen laut MEDLINE: Intubation, Intratracheal, Laryngoscopes. Die Arbeit nennt 22 Literaturangaben. Gefördert von: U.S. Department of Defense.
Abstract (Originalsprache)
Introduction
Video laryngoscope (VL) technology improves first‐pass success. The novel i‐view VL device is inexpensive and disposable. We sought to determine the first‐pass intubation success with the i‐view VL device versus the standard reusable VL systems in routine use at each site.
Methods
We performed a prospective, pragmatic study at two major emergency departments (EDs) when VL was used. We rotated i‐view versus reusable VL as the preferred device of the month based on an a priori schedule. An investigator‐initiated interim analysis was performed. Our primary outcome was a first‐pass success with a non‐inferiority margin of 10% based on the per‐protocol analysis.
Results
There were 93 intubations using the reusable VL devices and 81 intubations using the i‐view. Our study was stopped early due to futility in reaching our predetermined non‐inferiority margin. Operator and patient characteristics were similar between the two groups. The first‐pass success rate for the i‐view group was 69.1% compared to 84.3% for the reusable VL group. A non‐inferiority analysis indicated that the difference (−15.1%) and corresponding 90% confidence limits (−25.3% to −5.0%) did not fall within the predetermined 10% non‐inferiority margin.
Conclusions
The i‐view device failed to meet our predetermined non‐inferiority margin when compared to the reusable VL systems with the study stopping early due to futility. Significant crossover occurred at the discretion of the intubating operator during the i‐view month.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- US Army Institute of Surgical Research, JBSA Fort Sam Houston, Texas, USA.
- Brooke Army Medical Center, JBSA, JBSA Fort Sam Houston, Texas, USA.
- Uniformed Service University of the Health Sciences, Bethesda, Maryland, USA.
- 59th Medical Wing, JBSA Lackland, Texas, USA.
- 14th Field Hospital, Fort Stewart, Georgia, USA.
- University of Colorado School of Medicine, Aurora, Colorado, USA.
Förderung
- U.S. Department of Defense
Zitieren
APA
Schauer, S. G., Long, B. J., April, M. D., Resnick-Ault, D., Mendez, J., Arana, A. A., Bastman, J. J., Davis, W. T., Maddry, J. K., Ginde, A. A., & Bebarta, V. S. (2024). A prospective, pragmatic non‐inferiority study of emergency intubation success with the single‐use i‐view versus standard reusable video laryngoscope. Transfusion. https://doi.org/10.1111/trf.17790
Vancouver
Schauer SG, Long BJ, April MD, Resnick-Ault D, Mendez J, Arana AA, et al. A prospective, pragmatic non‐inferiority study of emergency intubation success with the single‐use i‐view versus standard reusable video laryngoscope. Transfusion. 2024. doi:10.1111/trf.17790
BibTeX
@article{Schauer2024,
author = {Schauer, Steven G and Long, Brit J and April, Michael D and Resnick-Ault, Daniel and Mendez, Jessica and Arana, Allyson A and Bastman, Jill J and Davis, William T and Maddry, Joseph K and Ginde, Adit A and Bebarta, Vikhyat S},
title = {A prospective, pragmatic non‐inferiority study of emergency intubation success with the single‐use i‐view versus standard reusable video laryngoscope},
journal = {Transfusion},
year = {2024},
doi = {10.1111/trf.17790}
}
EUVIMED