Fachartikel in der Zeitschrift Transfusion Medicine, online erschienen am 06.03.2026. Verfasst von Yohei Sakai und 6 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Health Sciences, Fujita Health University, Toyoake, Japan. Hauptthemen laut MEDLINE: Blood Preservation, Erythrocytes. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Ministry of Health, Labour and Welfare, Japan, Ministry of Health, Labour and Welfare.
Abstract (Originalsprache)
Background
Packed red blood cells (PRBC) are stored at 2°C–6°C to ensure quality. Improper temperature control during PRBC transport reduces the quality of downstream blood products and wastes PRBC units. This study evaluated the suitability of the BioBox LAB10 for in‐hospital PRBC transport.
Methods
Temperatures of the box interior and simulated formulation were measured to assess cooling capabilities. Quality was evaluated by measuring red blood cell count, haemoglobin concentration, haematocrit, pH, potassium concentration, and ATP concentration of PRBC samples. The storage capacity, size, weight, and cost of the BioBox was compared with that of the ATR700.
Results
The BioBox cooled to ≤6°C within 14 min. PRBC temperature remained ≤6°C for approximately 19 h. None of the quality parameters, including ATP concentration, differed significantly between samples stored in the BioBox or in a refrigerator. The BioBox is smaller, lighter, and 84% less expensive than the ATR700, with an equivalent storage capacity.
Conclusions
The BioBox effectively maintains temperature and PRBC quality during transport and provides a practical solution for in‐hospital transport of blood for transfusion owing to its compact, lightweight design, and affordability.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Graduate School of Health Sciences, Fujita Health University, Toyoake, Japan.
- Department of Blood Transfusion, Fujita Health University Hospital, Toyoake, Japan.
- Department of Cellular and Molecular Biology, Fujita Health University School of Medical Sciences, Toyoake, Japan.
- Department of Transfusion Medicine and Cell Therapy, Fujita Health University School of Medicine, Toyoake, Japan.
ORCID-Profile: Yohei Sakai, Hideaki Matsuura
Förderung
- Ministry of Health, Labour and Welfare, Japan
- Ministry of Health, Labour and Welfare
Zitieren
APA
Sakai, Y., Ozeki, T., Matsuura, H., Nakagawa, R., Ishihara, Y., Doi, H., & Miura, Y. (2026). Efficacy of the BioBox LAB10 for maintaining temperature and quality of packed red blood cells during in‐hospital storage and transport. Transfusion Medicine, 36(4), 406-410. https://doi.org/10.1111/tme.70073
Vancouver
Sakai Y, Ozeki T, Matsuura H, Nakagawa R, Ishihara Y, Doi H, et al. Efficacy of the BioBox LAB10 for maintaining temperature and quality of packed red blood cells during in‐hospital storage and transport. Transfusion Medicine. 2026;36(4):406-410. doi:10.1111/tme.70073
BibTeX
@article{Sakai2026,
author = {Sakai, Yohei and Ozeki, Takato and Matsuura, Hideaki and Nakagawa, Rie and Ishihara, Yuya and Doi, Hiroki and Miura, Yasuo},
title = {Efficacy of the BioBox LAB10 for maintaining temperature and quality of packed red blood cells during in‐hospital storage and transport},
journal = {Transfusion Medicine},
year = {2026},
volume = {36},
number = {4},
pages = {406--410},
doi = {10.1111/tme.70073}
}
EUVIMED