Effects of apparatus dead space on volumetric capnograms in neonates with healthy lungs: a simulation study

Marcelo D Campos, Lucio H Palazzi, Stephan H Böhm, Gerardo Tusman iD

Pediatric Anesthesia · 2023

DOI: 10.1111/pan.14724 ↗

At a glance

Journal article in Pediatric Anesthesia, 2023. Written by Marcelo D Campos and 3 co-authors. Institution: Department of Anesthesiology, Sanatorio Finochietto, Buenos Aires, Argentina. Major MEDLINE topics: Carbon Dioxide, Respiration, Artificial. 35 references.

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Background

Volumetric capnography in healthy ventilated neonates showed deformed waveforms, which are supposedly due to technological limitations of flow and carbon dioxide sensors.

Aims

This bench study analyzed the role of apparatus dead space on the shape of capnograms in simulated neonates with healthy lungs.

Methods

We simulated mechanical breaths in neonates of 2, 2.5, and 3 kg of body weight using a neonatal volumetric capnography simulator. The simulator was fed by a fixed amount of carbon dioxide of 6 mL/kg/min. Such simulator was ventilated in a volume control mode using fixed ventilatory settings with a tidal volume of 8 mL/kg and respiratory rates of 40, 35, and 30 breaths per minute for the 2, 2.5 and 3 kg neonates, respectively. We tested the above baseline ventilation with and without an additional apparatus dead space of 4 mL.

Results

Simulations showed that adding the apparatus dead space to baseline ventilation increased the amount of re‐inhaled carbon dioxide in all neonates: 0.16 ± 0.01 to 0.32 ± 0.03 mL (2 kg), 0.14 ± 0.02 to 0.39 ± 0.05 mL (2.5 kg), and 0.13 ± 0.01 to 0.36 ± 0.05 mL (3 kg); ( p < .001). Apparatus dead space was computed as part of the airway dead space, and therefore, the ratio of airway dead space to tidal volume increased from 0.51 ± 0.04 to 0.68 ± 0.06, from 0.43 ± 0.04 to 0.62 ± 0.01 and from 0.38 ± 0.01 to 0.60 ± 0.02 in the 2, 2.5 and 3 kg simulated neonates, respectively ( p < .001). Compared to baseline ventilation, adding apparatus dead space decreased the ratio of the volume of phase III to V T size from 31% to 11% (2 kg), from 40% to 16% (2.5 kg) and from 50% to 18% (3 kg); ( p < .001).

Conclusions

The addition of a small apparatus dead space artificially deformed the volumetric capnograms in simulated neonates with healthy lungs.

Subjects

MeSH terms (MEDLINE)

Carbon DioxideRespiration, ArtificialInfant, NewbornHumansRespiratory Dead SpaceLungTidal VolumeCapnography

Author keywords

airway managementcapnographymechanical ventilationnewbornrespiratory dead space

Substances

Carbon Dioxide

Authors and institutions

  • Department of Anesthesiology, Sanatorio Finochietto, Buenos Aires, Argentina.
  • Department of Anesthesiology, Children Hospital Dr. Orlando Alassia, Santa Fe, Argentina.
  • Clinic of Anesthesiology, Intensive Care Medicine and Pain Therapy, Rostock University Medical Center, Rostock, Germany.
  • Department of Anesthesia, Hospital Privado de Comunidad, Mar del Plata, Argentina.

ORCID profiles: Gerardo Tusman

Cite

APA

Campos, M. D., Palazzi, L. H., Böhm, S. H., & Tusman, G. (2023). Effects of apparatus dead space on volumetric capnograms in neonates with healthy lungs: a simulation study. Pediatric Anesthesia. https://doi.org/10.1111/pan.14724

Vancouver

Campos MD, Palazzi LH, Böhm SH, Tusman G. Effects of apparatus dead space on volumetric capnograms in neonates with healthy lungs: a simulation study. Pediatric Anesthesia. 2023. doi:10.1111/pan.14724

BibTeX
@article{Campos2023,
  author = {Campos, Marcelo D and Palazzi, Lucio H and Böhm, Stephan H and Tusman, Gerardo},
  title = {Effects of apparatus dead space on volumetric capnograms in neonates with healthy lungs: a simulation study},
  journal = {Pediatric Anesthesia},
  year = {2023},
  doi = {10.1111/pan.14724}
}
Start for free →