EUVIMEDTutustu EUVIMEDiin
Crossref

Contact Index for PulseSelect Catheter Apposition: Validation by Transesophageal Echocardiography

Gen Fujiwara, Yoshiaki Mizutani iD, Satoshi Yanagisawa iD, Toyoaki Murohara

Pacing and Clinical Electrophysiology · 2026-07-26

DOI: 10.1111/pace.70376 ↗

At a glance

Journal article in Pacing and Clinical Electrophysiology, published online on 2026-07-26. Written by Gen Fujiwara and 3 co-authors. Institution: Department of Cardiology, Yokkaichi Municipal Hospital, Yokkaichi, Mie, Japan. Cited 1 times so far (Crossref); 4 references.

TiivistelmäSubjectsBibliografiset tiedotCiteRelated studies

Abstract (original language)

Background

Reliable catheter–tissue apposition is important for effective pulmonary vein isolation using pulsed field ablation. The contact index, derived from real‐time impedance rise on the EnSite X mapping system, may provide an objective surrogate for catheter contact; however, its utility with the PulseSelect catheter remains insufficiently defined.

Methods

In this single‐center pilot study, patients with paroxysmal atrial fibrillation undergoing pulmonary vein isolation with the PulseSelect catheter under general anesthesia were evaluated. High‐resolution transesophageal echocardiography was used as the reference standard to classify electrode‐level catheter apposition as contact or non‐contact. Contact index values from electrodes 3–7 were paired with transesophageal echocardiography findings, and receiver operating characteristic curve analysis was performed.

Results

A total of 316 paired imaging–electrical data points were analyzed. Interobserver agreement for transesophageal echocardiography‐based contact assessment was excellent (Cohen's κ = 0.86). Contact index values were higher at contact than non‐contact segments. A contact index cutoff of 10% yielded an area under the curve of 0.864, with 76.1% sensitivity and 82.9% specificity. In the central electrodes 4–6, the area under the curve was 0.871, with 80.1% sensitivity and 81.2% specificity.

Conclusions

Contact index was associated with transesophageal echocardiography‐confirmed catheter–tissue apposition during pulsed field ablation using the PulseSelect catheter. A 10% threshold may be a useful candidate reference for real‐time apposition assessment, requiring validation in larger studies.

Subjects

Author keywords

atrial fibrillationcontact indexpulmonary vein isolationpulsed field ablationtransesophageal echocardiography

Authors and institutions

  • Department of Cardiology, Yokkaichi Municipal Hospital, Yokkaichi, Mie, Japan.
  • Department of Cardiology, Nagoya University Graduate School of Medicine, Nagoya, Aichi, Japan.

ORCID profiles: Yoshiaki Mizutani, Satoshi Yanagisawa

Cite

APA

Fujiwara, G., Mizutani, Y., Yanagisawa, S., & Murohara, T. (2026). Contact Index for PulseSelect Catheter Apposition: Validation by Transesophageal Echocardiography. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70376

Vancouver

Fujiwara G, Mizutani Y, Yanagisawa S, Murohara T. Contact Index for PulseSelect Catheter Apposition: Validation by Transesophageal Echocardiography. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70376

BibTeX
@article{Fujiwara2026,
  author = {Fujiwara, Gen and Mizutani, Yoshiaki and Yanagisawa, Satoshi and Murohara, Toyoaki},
  title = {Contact Index for PulseSelect Catheter Apposition: Validation by Transesophageal Echocardiography},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2026},
  doi = {10.1111/pace.70376}
}
Start for free →