EUVIMEDOpdag EUVIMED
Open access

Variability in Pacemaker Battery Estimates and Performance

Benjamin A Steinberg iD, Riley Gillette, Michael Torre, Prashanthan Sanders iD, Niraj Varma, Tina Baykaner, Thomas Deering iD, Andrea M Russo, Yue Zhang, Yaxu Zhuang, James V Freeman

Pacing and Clinical Electrophysiology · 2026-07-26

DOI: 10.1111/pace.70370 ↗

At a glance

Journal article in Pacing and Clinical Electrophysiology, published online on 2026-07-26. Written by Benjamin A Steinberg and 10 co-authors. Institution: Department of Medicine, Denver Health Medical Center, Denver, Colorado, USA. Cited 1 times so far (Crossref); 17 references. The full text is freely available in PubMed Central. Funded by: NIH/NHLBI, NHLBI NIH HHS.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Permanent pacemaker (PPM) battery longevity impacts downstream procedures, patient outcomes, and healthcare costs.

Objectives

To measure and predict PPM battery longevity in routine, clinical practice.

Methods

We analyzed a nationwide, multicenter, remote monitoring dataset (PaceMate) to evaluate PPM battery longevity from 2007 to 2023. Multivariable models identified factors associated with replacement interval (RI) across device types and manufacturers. We also compared observed versus expected battery longevity and assessed differences between newer and older generation pacemakers.

Results

We included 58,395 devices from 47 sites: 5803 single‐chamber transvenous, 1637 single‐chamber leadless, 45,293 dual‐chamber transvenous, 1759 RV/LV‐only CRT pacemakers, and 3903 full CRT‐P systems (1888 reaching RI). Median observed battery longevity for the 1493 dual chamber transvenous pacemakers that reached RI ranged from (103 months for BSX to 125 months for MDT and BIO), with similar trends after adjustment ( p

Subjects

Author keywords

battery longevitymanufacturerpermanent pacemaker

Authors and institutions

  • Department of Medicine, Denver Health Medical Center, Denver, Colorado, USA.
  • Department of Medicine, University of Colorado Anschutz Medical Campus, Aurora, Colorado, USA.
  • CPC Clinical Research, Aurora, Colorado, USA.
  • Department of Medicine, University of Utah Health Sciences Center, Salt Lake City, Utah, USA.
  • Centre For Heart Rhythm Disorders (CHRD), University of Adelaide and Royal Adelaide, Hospital, Adelaide, Australia.
  • Department of Medicine, Cleveland Clinic, Cleveland, Ohio, USA.

ORCID profiles: Benjamin A Steinberg, Prashanthan Sanders, Thomas Deering

Funding

  • NIH/NHLBI
  • NHLBI NIH HHS

Cite

APA

Steinberg, B. A., Gillette, R., Torre, M., Sanders, P., Varma, N., Baykaner, T., Deering, T., Russo, A. M., Zhang, Y., Zhuang, Y., & Freeman, J. V. (2026). Variability in Pacemaker Battery Estimates and Performance. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70370

Vancouver

Steinberg BA, Gillette R, Torre M, Sanders P, Varma N, Baykaner T, et al. Variability in Pacemaker Battery Estimates and Performance. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70370

BibTeX
@article{Steinberg2026,
  author = {Steinberg, Benjamin A and Gillette, Riley and Torre, Michael and Sanders, Prashanthan and Varma, Niraj and Baykaner, Tina and Deering, Thomas and Russo, Andrea M and Zhang, Yue and Zhuang, Yaxu and Freeman, James V},
  title = {Variability in Pacemaker Battery Estimates and Performance},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2026},
  doi = {10.1111/pace.70370}
}
Start for free →