Comparison of Ablation Index‐Guided Lesion Size in nGEN and SMARTABLATE: An Ex Vivo Porcine Heart Study

Xin-Ru Li iD, Jie-Wen Yin, Xue-Si Wang, Yang-Yang Feng, Jia-Lin Li, Xiao-Ying Liu iD, Zhuo-Hang Du, Wen-He Lv, Tong Liu iD, Meng-Meng Li, Biao Fu, Chang-Qi Jia and 10 more

Pacing and Clinical Electrophysiology · 2026-07-09

DOI: 10.1111/pace.70353 ↗

At a glance

Journal article in Pacing and Clinical Electrophysiology, published online on 2026-07-09. Written by Xin-Ru Li and 21 co-authors. Institution: Department of Cardiology, Beijing Anzhen Hospital, Beijing Institute of Heart Lung and Blood Vessel Diseases, Capital Medical University, Beijing, Ch…. 16 references. Funded by: National Natural Science Foundation of China, Zhongnanshan Medical Foundation of Guangdong Province, Beijing Xinlian Zhicheng Cardiovascular Health Public Welfare Foundation.

TiivistelmäSubjectsBibliografiset tiedotCiteRelated studies

Abstract (original language)

Background nGEN TM generator is a new‐generation radiofrequency (RF) energy device widely used in the catheter ablation of atrial fibrillation. Compared with SMARTABLATE TM (SMA), a lower displayed impedance of nGEN can be observed during electrophysiology procedures. However, whether this impedance disparity affects ablation lesion size remains unelucidated.

Methods

Fresh ex vivo porcine hearts were used for the experiment. Impedance of the same marked points on porcine left ventricle was measured using both generators. Ablations were performed at varying RF power (35 W, 45 W) and ablation index (AI; 400, 450, 550). Five lesions for each combination of settings were created with both generators and their lesion sizes were compared.

Results

The impedance of nGEN was notably lower than SMA (78.1±3.2Ω vs. 87.5±4.2Ω, p < 0.05). Under identical application settings, lesion depth, maximum length, and volume did not differ significantly between the two generators in any of the groups. Surface measurements of maximum diameter and area also showed no significant differences except at 45 W/AI = 450, where nGEN generated significantly larger values than SMA (6.5 [6.4–6.9]mm vs. 5.8 [5.5–6.0]mm; 22.8 [22.5–24.4]mm 2 vs. 20.3 [18.8–20.5]mm 2 ; all p < 0.05). Lesion depth, maximum length, and volume of both generators were strongly correlated with AI (nGEN: r = 0.796, r = 0.657, r = 0.769; SMA: r = 0.794, r = 0.678, r = 0.733; all p < 0.05). Conclusions nGEN displays lower impedance than SMA. Lesion sizes are overall comparable between the two generators in AI‐guided ablation, with the exception of surface measurements at 45 W/AI = 450.

Subjects

Author keywords

SMAablation lesionimpedancenGENradiofrequency ablation

Authors and institutions

  • Department of Cardiology, Beijing Anzhen Hospital, Beijing Institute of Heart Lung and Blood Vessel Diseases, Capital Medical University, Beijing, China.

ORCID profiles: Xin-Ru Li, Xiao-Ying Liu, Tong Liu, De-Yong Long, Chang-Sheng Ma

Funding

  • National Natural Science Foundation of China
  • Zhongnanshan Medical Foundation of Guangdong Province
  • Beijing Xinlian Zhicheng Cardiovascular Health Public Welfare Foundation

Cite

APA

Li, X. R., Yin, J. W., Wang, X. S., Feng, Y. Y., Li, J. L., Liu, X. Y., Du, Z. H., Lv, W. H., Liu, T., Li, M. M., Fu, B., Jia, C. Q., Wang, W., Wu, J. H., Zhao, X., Li, C. Y., Sang, C. H., Jiang, C. X., Long, D. Y., …, & Tang, R. B. (2026). Comparison of Ablation Index‐Guided Lesion Size in nGEN and SMARTABLATE: An Ex Vivo Porcine Heart Study. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70353

Vancouver

Li XR, Yin JW, Wang XS, Feng YY, Li JL, Liu XY, et al. Comparison of Ablation Index‐Guided Lesion Size in nGEN and SMARTABLATE: An Ex Vivo Porcine Heart Study. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70353

BibTeX
@article{Li2026,
  author = {Li, Xin-Ru and Yin, Jie-Wen and Wang, Xue-Si and Feng, Yang-Yang and Li, Jia-Lin and Liu, Xiao-Ying and Du, Zhuo-Hang and Lv, Wen-He and Liu, Tong and Li, Meng-Meng and Fu, Biao and Jia, Chang-Qi and Wang, Wei and Wu, Jia-Hui and Zhao, Xin and Li, Chang-Yi and Sang, Cai-Hua and Jiang, Chen-Xi and Long, De-Yong and Dong, Jian-Zeng and Ma, Chang-Sheng and Tang, Ri-Bo},
  title = {Comparison of Ablation Index‐Guided Lesion Size in nGEN and SMARTABLATE: An Ex Vivo Porcine Heart Study},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2026},
  doi = {10.1111/pace.70353}
}
Start for free →