Modular Upgrade From AVEIR VR to AVEIR DR: A Leadless Dual‐Chamber Pacing Experience in Central Asia
Pacing and Clinical Electrophysiology · 2026-06 · 49 (6) · 808-810
DOI: 10.1111/pace.70212 ↗Kurz gefasst
Fallbericht in der Zeitschrift Pacing and Clinical Electrophysiology, online erschienen am 19.03.2026. Verfasst von Ayan Abdrakhmanov und 3 weiteren Autor:innen. Beteiligte Einrichtung: Heart Rhythm Scientific Research Institute, NpJSC "Astana Medical University", Astana, Kazakhstan. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Atrioventricular Block, Cardiac Pacing, Artificial, Sick Sinus Syndrome. Die Arbeit nennt 3 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Leadless pacing systems have evolved toward modular configurations that enable staged restoration of atrioventricular (AV) synchrony without transvenous leads. Real‐world experience with a sequential upgrade from AVEIR VR to a dual‐chamber AVEIR DR system remains limited. We report a case of staged modular leadless upgrade performed in a patient with progressive sinus node dysfunction, intermittent high‐grade AV block, and symptomatic loss of AV synchrony following prior ventricular leadless pacing. A 74‐year‐old man with a history of mitral valve repair and atrial flutter ablation initially received an AVEIR VR ventricular device. One year later, he developed sinus pauses and hemodynamic symptoms related to AV dyssynchrony. Because the patient declined transvenous pacing, an atrial AVEIR AR module was implanted to establish dual‐chamber pacing within a fully leadless system. The procedure was uncomplicated and resulted in restoration of AV synchrony and clinical improvement at follow‐up. This report demonstrates the feasibility of staged modular upgrades within a leadless pacing platform in selected patients with evolving conduction disease.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Heart Rhythm Scientific Research Institute, NpJSC "Astana Medical University", Astana, Kazakhstan.
- Heart Rhythm Center, National Coordination Center for Emergency Medicine, Ministry of Healthcare of the Republic of Kazakhstan, Astana, Kazakhstan.
ORCID-Profile: Gaziza Marat
Zitieren
APA
Abdrakhmanov, A., Marat, G., Yessimbekova, E., & Suleymen, Z. (2026). Modular Upgrade From AVEIR VR to AVEIR DR: A Leadless Dual‐Chamber Pacing Experience in Central Asia. Pacing and Clinical Electrophysiology, 49(6), 808-810. https://doi.org/10.1111/pace.70212
Vancouver
Abdrakhmanov A, Marat G, Yessimbekova E, Suleymen Z. Modular Upgrade From AVEIR VR to AVEIR DR: A Leadless Dual‐Chamber Pacing Experience in Central Asia. Pacing and Clinical Electrophysiology. 2026;49(6):808-810. doi:10.1111/pace.70212
BibTeX
@article{Abdrakhmanov2026,
author = {Abdrakhmanov, Ayan and Marat, Gaziza and Yessimbekova, Elmira and Suleymen, Zhanasyl},
title = {Modular Upgrade From AVEIR VR to AVEIR DR: A Leadless Dual‐Chamber Pacing Experience in Central Asia},
journal = {Pacing and Clinical Electrophysiology},
year = {2026},
volume = {49},
number = {6},
pages = {808--810},
doi = {10.1111/pace.70212}
}
EUVIMED