Safety Outcomes of Hybrid Open Chest Transvenous Lead Extraction: A Multicenter Experience
Pacing and Clinical Electrophysiology · 2025
DOI: 10.1111/pace.70118 ↗At a glance
Multizentrische Studie in Pacing and Clinical Electrophysiology, 2025. Written by Nadeev Wijesuriya and 15 co-authors. Institution: King's College London, London, UK. Major MEDLINE topics: Device Removal, Pacemaker, Artificial, Defibrillators, Implantable. 28 references. The full text is freely available (open access, CC BY 4.0). Funded by: European Research Council, Centre For Medical Engineering, King's College London, British Heart Foundation.
Abstract (original language)
Background
Hybrid open chest transvenous lead extraction (TLE), combining surgical and endovascular techniques, may be utilized in patients requiring concomitant cardiac surgery or with high‐risk features for endovascular extraction. Outcome data in this population remains sparse.
Aims
To evaluate procedural outcomes and identify predictors of complications in patients undergoing elective hybrid open chest TLE.
Methods
A retrospective multicenter cohort study was conducted, including 40 patients between 2017 and 2025 across three UK tertiary centers. Patients undergoing emergency surgical conversion were excluded. Baseline, procedural, and outcome data were collected. Primary outcomes were in‐hospital mortality and complications, graded using a modified Delphi classification. Logistic regression was used to identify predictors of any or severe (Delphi grade ≥3) complications.
Results
The mean age was 61.9 ± 17 years; 62.5% were male. Median lead dwell time was 10.5 years. The indication was infection in 65% of cases. Concomitant valve intervention was performed in 77.5% of cases. Clinical procedural success was 97.5%. In‐hospital mortality was 2.5%, with a rate of severe complications of 15%, and an overall complication rate of 37.5%. Multivariable analysis identified chronic lung disease as an independent predictor of severe complications (OR 102.2, p = 0.03). Atrial fibrillation was an independent predictor of any complication (OR 5.83, p = 0.04), driven primarily by post‐procedure rhythm intervention. Lead dwell time and EROS classification were not associated with adverse outcomes.
Conclusion
Hybrid open chest TLE demonstrates high procedural success and despite significant morbidity, has acceptable mortality rates. Chronic lung disease independently predicts complications and should guide patient selection and perioperative planning.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- King's College London, London, UK.
- Guy's and St Thomas's NHS Foundation Trust, London, UK.
- Imperial College London, London, UK.
- Edinburgh Heart Centre, Edinburgh, UK.
- Alan Turing Institute, London, UK.
- Cleveland Clinic, London, UK.
ORCID profiles: Nadeev Wijesuriya, Alphonsus Liew, Steven A Niederer
Funding
- European Research Council
- Centre For Medical Engineering, King's College London
- British Heart Foundation
Cite
APA
Wijesuriya, N., Cobbold, H. L., Kellman, K., Vogiatzakis, N., Vere, F., Howell, S., Liew, A., Behar, J. M., Bosco, P., Lucchese, G., Markides, V., Grubb, N. R., Mehta, V., Niederer, S. A., Wong, T., & Rinaldi, C. A. (2025). Safety Outcomes of Hybrid Open Chest Transvenous Lead Extraction: A Multicenter Experience. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70118
Vancouver
Wijesuriya N, Cobbold HL, Kellman K, Vogiatzakis N, Vere F, Howell S, et al. Safety Outcomes of Hybrid Open Chest Transvenous Lead Extraction: A Multicenter Experience. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.70118
BibTeX
@article{Wijesuriya2025,
author = {Wijesuriya, Nadeev and Cobbold, Helena Lytton and Kellman, Keisha and Vogiatzakis, Nikolaos and Vere, Felicity De and Howell, Sandra and Liew, Alphonsus and Behar, Jonathan M and Bosco, Paolo and Lucchese, Gianluca and Markides, Vias and Grubb, Neil R and Mehta, Vishal and Niederer, Steven A and Wong, Tom and Rinaldi, Christopher A},
title = {Safety Outcomes of Hybrid Open Chest Transvenous Lead Extraction: A Multicenter Experience},
journal = {Pacing and Clinical Electrophysiology},
year = {2025},
doi = {10.1111/pace.70118}
}
EUVIMED