Safety Outcomes of Hybrid Open Chest Transvenous Lead Extraction: A Multicenter Experience

Nadeev Wijesuriya iD, Helena Lytton Cobbold, Keisha Kellman, Nikolaos Vogiatzakis, Felicity De Vere, Sandra Howell, Alphonsus Liew iD, Jonathan M Behar, Paolo Bosco, Gianluca Lucchese, Vias Markides, Neil R Grubb und 4 weitere

Pacing and Clinical Electrophysiology · 2025

DOI: 10.1111/pace.70118 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Pacing and Clinical Electrophysiology, 2025. Verfasst von Nadeev Wijesuriya und 15 weiteren Autor:innen. Beteiligte Einrichtung: King's College London, London, UK. Hauptthemen laut MEDLINE: Device Removal, Pacemaker, Artificial, Defibrillators, Implantable. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Research Council, Centre For Medical Engineering, King's College London, British Heart Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Hybrid open chest transvenous lead extraction (TLE), combining surgical and endovascular techniques, may be utilized in patients requiring concomitant cardiac surgery or with high‐risk features for endovascular extraction. Outcome data in this population remains sparse.

Aims

To evaluate procedural outcomes and identify predictors of complications in patients undergoing elective hybrid open chest TLE.

Methods

A retrospective multicenter cohort study was conducted, including 40 patients between 2017 and 2025 across three UK tertiary centers. Patients undergoing emergency surgical conversion were excluded. Baseline, procedural, and outcome data were collected. Primary outcomes were in‐hospital mortality and complications, graded using a modified Delphi classification. Logistic regression was used to identify predictors of any or severe (Delphi grade ≥3) complications.

Results

The mean age was 61.9 ± 17 years; 62.5% were male. Median lead dwell time was 10.5 years. The indication was infection in 65% of cases. Concomitant valve intervention was performed in 77.5% of cases. Clinical procedural success was 97.5%. In‐hospital mortality was 2.5%, with a rate of severe complications of 15%, and an overall complication rate of 37.5%. Multivariable analysis identified chronic lung disease as an independent predictor of severe complications (OR 102.2, p = 0.03). Atrial fibrillation was an independent predictor of any complication (OR 5.83, p = 0.04), driven primarily by post‐procedure rhythm intervention. Lead dwell time and EROS classification were not associated with adverse outcomes.

Conclusion

Hybrid open chest TLE demonstrates high procedural success and despite significant morbidity, has acceptable mortality rates. Chronic lung disease independently predicts complications and should guide patient selection and perioperative planning.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Device RemovalPacemaker, ArtificialDefibrillators, ImplantableHumansMaleFemaleMiddle AgedRetrospective StudiesHospital MortalityPostoperative ComplicationsAgedDevice Lead Extraction

Stichwörter der Autor:innen

cardiac implantable electronic devicecardiac surgeryextraction

Autor:innen und Institutionen

  • King's College London, London, UK.
  • Guy's and St Thomas's NHS Foundation Trust, London, UK.
  • Imperial College London, London, UK.
  • Edinburgh Heart Centre, Edinburgh, UK.
  • Alan Turing Institute, London, UK.
  • Cleveland Clinic, London, UK.

ORCID-Profile: Nadeev Wijesuriya, Alphonsus Liew, Steven A Niederer

Förderung

  • European Research Council
  • Centre For Medical Engineering, King's College London
  • British Heart Foundation

Zitieren

APA

Wijesuriya, N., Cobbold, H. L., Kellman, K., Vogiatzakis, N., Vere, F., Howell, S., Liew, A., Behar, J. M., Bosco, P., Lucchese, G., Markides, V., Grubb, N. R., Mehta, V., Niederer, S. A., Wong, T., & Rinaldi, C. A. (2025). Safety Outcomes of Hybrid Open Chest Transvenous Lead Extraction: A Multicenter Experience. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70118

Vancouver

Wijesuriya N, Cobbold HL, Kellman K, Vogiatzakis N, Vere F, Howell S, et al. Safety Outcomes of Hybrid Open Chest Transvenous Lead Extraction: A Multicenter Experience. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.70118

BibTeX
@article{Wijesuriya2025,
  author = {Wijesuriya, Nadeev and Cobbold, Helena Lytton and Kellman, Keisha and Vogiatzakis, Nikolaos and Vere, Felicity De and Howell, Sandra and Liew, Alphonsus and Behar, Jonathan M and Bosco, Paolo and Lucchese, Gianluca and Markides, Vias and Grubb, Neil R and Mehta, Vishal and Niederer, Steven A and Wong, Tom and Rinaldi, Christopher A},
  title = {Safety Outcomes of Hybrid Open Chest Transvenous Lead Extraction: A Multicenter Experience},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2025},
  doi = {10.1111/pace.70118}
}
Kostenlos starten →