Horizontal Axis Affects Paced QRS Duration More Than Vertical Axis at Leadless Pacemaker Implantation Site
Pacing and Clinical Electrophysiology · 2025
DOI: 10.1111/pace.70107 ↗Kurz gefasst
Fachartikel in der Zeitschrift Pacing and Clinical Electrophysiology, 2025. Verfasst von Hirotsugu Sato und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Dokkyo Medical University Saitama Medical Center, Saitama, Japan. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Cardiomyopathies, Prosthesis Implantation, Cardiac Pacing, Artificial. Die Arbeit nennt 18 Literaturangaben.
Abstract (Originalsprache)
Background
Leadless pacemaker (LP) implantation at the apical septum has been linked to pacing‐induced cardiomyopathy (PICM). This study aimed to examine the precise position for LP implantation in the RV septum using the three‐dimensional rotational angiography (3D‐RA) system. The accurate implantation sites were visualized through the fluoroscopic image, and the relationship to the measured QRS duration during pacing (p‐QRS duration) was investigated to evaluate their potential association with cardiomyopathy.
Methods
Fifty‐four patients underwent LP implantation with 3D‐RA guidance. Implantation sites were assessed using the horizontal axis from the apex to the center of the tricuspid valve and the vertical axis from the inferior border of the apex to the outflow tract transition. The p‐QRS durations were measured intraoperatively using electrocardiograms. Echocardiography was performed 12 months later to assess PCIM progression.
Results
All 54 LPs were successfully implanted in the septum. Mean horizontal and vertical positions within the RV were 69.7 ± 10.5% and 66.5 ± 13.1%, respectively. Mean p‐QRS duration was 148 ± 11 ms. The horizontal position was correlated with QRS duration ( R = 0.59), while the vertical position showed no correlation ( R = 0.03). The native QRS (n‐QRS) duration was also an independent predictor of prolonged p‐QRS duration ( p = 0.018). No patients developed cardiomyopathy progression on follow‐up, and no association with p‐QRS duration was observed.
Conclusion
In LP implantation, the horizontal position within the RV septum correlates more strongly with p‐QRS duration than vertical position. Prolonged n‐QRS duration also prolonged p‐QRS duration; however, further research is needed to evaluate the relationship between the implanted position and the occurrence of PICM.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Dokkyo Medical University Saitama Medical Center, Saitama, Japan.
- Department of Cardiology, Meirikai Chuo General Hospital, Tokyo, Japan.
ORCID-Profile: Yuichi Hori
Zitieren
APA
Sato, H., Hori, Y., Kimura, Y., Aoki, H., Fukuda, R., Hisauchi, I., Nakahara, S., Itabashi, Y., Ishikawa, T., Kobayashi, S., & Taguchi, I. (2025). Horizontal Axis Affects Paced QRS Duration More Than Vertical Axis at Leadless Pacemaker Implantation Site. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70107
Vancouver
Sato H, Hori Y, Kimura Y, Aoki H, Fukuda R, Hisauchi I, et al. Horizontal Axis Affects Paced QRS Duration More Than Vertical Axis at Leadless Pacemaker Implantation Site. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.70107
BibTeX
@article{Sato2025,
author = {Sato, Hirotsugu and Hori, Yuichi and Kimura, Yuta and Aoki, Hideyuki and Fukuda, Reiko and Hisauchi, Itaru and Nakahara, Shiro and Itabashi, Yuji and Ishikawa, Tetsuya and Kobayashi, Sayuki and Taguchi, Isao},
title = {Horizontal Axis Affects Paced QRS Duration More Than Vertical Axis at Leadless Pacemaker Implantation Site},
journal = {Pacing and Clinical Electrophysiology},
year = {2025},
doi = {10.1111/pace.70107}
}
EUVIMED