Ventricular Tachycardia Due to Blanking of a Premature Ventricular Contraction
Pacing and Clinical Electrophysiology · 2025
DOI: 10.1111/pace.70075 ↗At a glance
Fallbericht in Pacing and Clinical Electrophysiology, 2025. Written by Shing Ching and 1 co-authors. Institution: Division of Cardiology, Department of Medicine and Geriatrics, United Christian Hospital, Hong Kong, China. Major MEDLINE topics: Tachycardia, Ventricular, Ventricular Premature Complexes, Pacemaker, Artificial, Cardiac Pacing, Artificial. 16 references.
Abstract (original language)
A 78‐year‐old man was evaluated for a non‐sustained ventricular tachycardia episode detected in a dual‐chamber pacemaker (Boston Scientific Essentio MRI EL L131) implanted for sick sinus syndrome. Blanking of a premature ventricular beat caused delivery of ventricular pacing in the vulnerable period triggering non‐sustained ventricular tachycardia. Review of literature found more than half of reported cases involved Boston devices. The commonalities of such phenomenon are summarized and reprogramming strategies presented.
Subjects
MeSH terms (MEDLINE)
Authors and institutions
- Division of Cardiology, Department of Medicine and Geriatrics, United Christian Hospital, Hong Kong, China.
ORCID profiles: Shing Ching
Cite
APA
Ching, S., & Lau, A. T. (2025). Ventricular Tachycardia Due to Blanking of a Premature Ventricular Contraction. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70075
Vancouver
Ching S, Lau AT. Ventricular Tachycardia Due to Blanking of a Premature Ventricular Contraction. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.70075
BibTeX
@article{Ching2025,
author = {Ching, Shing and Lau, Alexson Tsz-Ki},
title = {Ventricular Tachycardia Due to Blanking of a Premature Ventricular Contraction},
journal = {Pacing and Clinical Electrophysiology},
year = {2025},
doi = {10.1111/pace.70075}
}
EUVIMED