Fallbericht in der Zeitschrift Pacing and Clinical Electrophysiology, 2025. Verfasst von Shing Ching und 1 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Department of Medicine and Geriatrics, United Christian Hospital, Hong Kong, China. Hauptthemen laut MEDLINE: Tachycardia, Ventricular, Ventricular Premature Complexes, Pacemaker, Artificial, Cardiac Pacing, Artificial. Die Arbeit nennt 16 Literaturangaben.
Abstract (Originalsprache)
A 78‐year‐old man was evaluated for a non‐sustained ventricular tachycardia episode detected in a dual‐chamber pacemaker (Boston Scientific Essentio MRI EL L131) implanted for sick sinus syndrome. Blanking of a premature ventricular beat caused delivery of ventricular pacing in the vulnerable period triggering non‐sustained ventricular tachycardia. Review of literature found more than half of reported cases involved Boston devices. The commonalities of such phenomenon are summarized and reprogramming strategies presented.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Division of Cardiology, Department of Medicine and Geriatrics, United Christian Hospital, Hong Kong, China.
ORCID-Profile: Shing Ching
Zitieren
APA
Ching, S., & Lau, A. T. (2025). Ventricular Tachycardia Due to Blanking of a Premature Ventricular Contraction. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70075
Vancouver
Ching S, Lau AT. Ventricular Tachycardia Due to Blanking of a Premature Ventricular Contraction. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.70075
BibTeX
@article{Ching2025,
author = {Ching, Shing and Lau, Alexson Tsz-Ki},
title = {Ventricular Tachycardia Due to Blanking of a Premature Ventricular Contraction},
journal = {Pacing and Clinical Electrophysiology},
year = {2025},
doi = {10.1111/pace.70075}
}
EUVIMED