EUVIMEDScopri EUVIMED
Open access · CC BY-NC 4.0

Electromagnetic force from electric vehicles: Potential electromagnetic interference source for subcutaneous implantable defibrillator

Anmar Salih iD, Kamil Salam, Tarun Goswami, Abdul Wase

Pacing and Clinical Electrophysiology · 2024-07 · 47 (7) · 885-892

DOI: 10.1111/pace.15019 ↗

At a glance

Journal article in Pacing and Clinical Electrophysiology, published online on 2024-06-03. Written by Anmar Salih and 3 co-authors. Institution: Department of Biomedical, Industrial and Human Factors Engineering, Wright State University, Dayton, USA. Major MEDLINE topics: Defibrillators, Implantable, Electromagnetic Fields. Cited 1 times so far (Crossref); 19 references. The full text is freely available (open access, CC BY-NC 4.0).

AbstractSubjectsDati bibliograficiCiteRelated studies

Abstract (original language)

Background

Electromagnetic interference (EMI) encompasses electromagnetic field signals that can be detected by a device's circuitry, potentially resulting in adverse effects such as inaccurate sensing, pacing, device mode switching, and defibrillation. EMI may impact the functioning of Cardiac Implantable Electronic Devices (CIEDs) and lead to inappropriate therapy.

Method

An experimental measuring device, a loop antenna mimicking the implantable cardioverted defibrillator (ICD) antenna, was developed, and validated at the US Food and Drug Administration (FDA) and sent to Wright State University for testing. Two sets of measurements were conducted while the vehicle was connected to a 220‐Volt outlet with charging at ON and OFF. Each measurement set involved three readings at various locations, with the antenna oriented in three different positions to account for diverse patient postures. The experiment utilized a Tesla Model 3 electric vehicle (EV), assessing scenarios both inside and outside the car, including the driver's seat, driver's seat floor, passenger's seat, rear seat, rear seat floor, cup holder, charging port (car), and near the charging station.

Results

The detected voltage (max 400 to 504 millivolts) around the cup holder inside the car differed from all other measurement scenarios.

Conclusion

The investigation highlights the identification of EMI signals originating from an EV) that could potentially interrupt the functionality of a Subcutaneous Implantable Cardioverter‐Defibrillator (S‐ICD). These signals fell within the R‐wave Spectrum of 30–300 Hz. Further in‐vivo studies are essential to determine accurately the level of interference between S‐ICDs and EMI from Electric Vehicles.

Subjects

MeSH terms (MEDLINE)

Defibrillators, ImplantableElectromagnetic FieldsHumansEquipment Failure AnalysisAutomobiles

Author keywords

EMFEMIEVS‐ICDTesla Model 3electric vehiclesubcutaneous ICD

Authors and institutions

  • Department of Biomedical, Industrial and Human Factors Engineering, Wright State University, Dayton, USA.
  • Wright State University Boonshoft School of Medicine, Internal Medicine, Dayton, USA.
  • Premier Cardiovascular Institute, Wright State University Boonshoft School of Medicine, Premier Health, Dayton, USA.

ORCID profiles: Anmar Salih

Cite

APA

Salih, A., Salam, K., Goswami, T., & Wase, A. (2024). Electromagnetic force from electric vehicles: Potential electromagnetic interference source for subcutaneous implantable defibrillator. Pacing and Clinical Electrophysiology, 47(7), 885-892. https://doi.org/10.1111/pace.15019

Vancouver

Salih A, Salam K, Goswami T, Wase A. Electromagnetic force from electric vehicles: Potential electromagnetic interference source for subcutaneous implantable defibrillator. Pacing and Clinical Electrophysiology. 2024;47(7):885-892. doi:10.1111/pace.15019

BibTeX
@article{Salih2024,
  author = {Salih, Anmar and Salam, Kamil and Goswami, Tarun and Wase, Abdul},
  title = {Electromagnetic force from electric vehicles: Potential electromagnetic interference source for subcutaneous implantable defibrillator},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2024},
  volume = {47},
  number = {7},
  pages = {885--892},
  doi = {10.1111/pace.15019}
}