Electromagnetic force from electric vehicles: Potential electromagnetic interference source for subcutaneous implantable defibrillator
Pacing and Clinical Electrophysiology · 2024-07 · 47 (7) · 885-892
DOI: 10.1111/pace.15019 ↗At a glance
Journal article in Pacing and Clinical Electrophysiology, published online on 2024-06-03. Written by Anmar Salih and 3 co-authors. Institution: Department of Biomedical, Industrial and Human Factors Engineering, Wright State University, Dayton, USA. Major MEDLINE topics: Defibrillators, Implantable, Electromagnetic Fields. Cited 1 times so far (Crossref); 19 references. The full text is freely available (open access, CC BY-NC 4.0).
Abstract (original language)
Background
Electromagnetic interference (EMI) encompasses electromagnetic field signals that can be detected by a device's circuitry, potentially resulting in adverse effects such as inaccurate sensing, pacing, device mode switching, and defibrillation. EMI may impact the functioning of Cardiac Implantable Electronic Devices (CIEDs) and lead to inappropriate therapy.
Method
An experimental measuring device, a loop antenna mimicking the implantable cardioverted defibrillator (ICD) antenna, was developed, and validated at the US Food and Drug Administration (FDA) and sent to Wright State University for testing. Two sets of measurements were conducted while the vehicle was connected to a 220‐Volt outlet with charging at ON and OFF. Each measurement set involved three readings at various locations, with the antenna oriented in three different positions to account for diverse patient postures. The experiment utilized a Tesla Model 3 electric vehicle (EV), assessing scenarios both inside and outside the car, including the driver's seat, driver's seat floor, passenger's seat, rear seat, rear seat floor, cup holder, charging port (car), and near the charging station.
Results
The detected voltage (max 400 to 504 millivolts) around the cup holder inside the car differed from all other measurement scenarios.
Conclusion
The investigation highlights the identification of EMI signals originating from an EV) that could potentially interrupt the functionality of a Subcutaneous Implantable Cardioverter‐Defibrillator (S‐ICD). These signals fell within the R‐wave Spectrum of 30–300 Hz. Further in‐vivo studies are essential to determine accurately the level of interference between S‐ICDs and EMI from Electric Vehicles.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Biomedical, Industrial and Human Factors Engineering, Wright State University, Dayton, USA.
- Wright State University Boonshoft School of Medicine, Internal Medicine, Dayton, USA.
- Premier Cardiovascular Institute, Wright State University Boonshoft School of Medicine, Premier Health, Dayton, USA.
ORCID profiles: Anmar Salih
Cite
APA
Salih, A., Salam, K., Goswami, T., & Wase, A. (2024). Electromagnetic force from electric vehicles: Potential electromagnetic interference source for subcutaneous implantable defibrillator. Pacing and Clinical Electrophysiology, 47(7), 885-892. https://doi.org/10.1111/pace.15019
Vancouver
Salih A, Salam K, Goswami T, Wase A. Electromagnetic force from electric vehicles: Potential electromagnetic interference source for subcutaneous implantable defibrillator. Pacing and Clinical Electrophysiology. 2024;47(7):885-892. doi:10.1111/pace.15019
BibTeX
@article{Salih2024,
author = {Salih, Anmar and Salam, Kamil and Goswami, Tarun and Wase, Abdul},
title = {Electromagnetic force from electric vehicles: Potential electromagnetic interference source for subcutaneous implantable defibrillator},
journal = {Pacing and Clinical Electrophysiology},
year = {2024},
volume = {47},
number = {7},
pages = {885--892},
doi = {10.1111/pace.15019}
}
EUVIMED