Bailing Dai, Wenqing Yin, Chengming Ma, Peipei Ma iD, Yingqi Wang, Li Hong, Xiaohong Yu, Xiaomeng Yin, Yingxue Dong, Yunlong Xia, Yanzong Yang, Lianjun Gao iD

Pacing and Clinical Electrophysiology · 2024

DOI: 10.1111/pace.14993 ↗

At a glance

Fallbericht in Pacing and Clinical Electrophysiology, 2024. Written by Bailing Dai and 11 co-authors. Institution: Department of cardiology, The First Affiliated Hospital of Dalian Medical University, Dalian, China. Major MEDLINE topics: Equipment Failure, Pacemaker, Artificial. 4 references.

RezumatSubjectsDate bibliograficeCiteRelated studies

Abstract (original language)

Although ventricular capture during the atrial threshold test is possible, there are rare reports on the insulation defect and inactive leads thereof. In this case, we present a pacemaker‐dependent patient with a history of pacemaker generator replacements. The patient experienced ventricular capture induced by atrial pacing due to adhesion of the atrial and ventricular leads with an insulation defect. The atrial lead was abandoned and a new lead was implanted. However, there was a significant decrease in ventricular impedance detected shortly after the new lead was implanted. When observing the phenomenon of atrial pacing‐induced ventricular depolarization, one uncommon reason to consider is lead adhesive wear. It is important to pay attention to the contact and bending sites of the leads.

Subjects

MeSH terms (MEDLINE)

Equipment FailurePacemaker, ArtificialHumansElectrodes, ImplantedMaleCardiac Pacing, ArtificialAgedHeart Atria

Author keywords

insulation defectlead adhesive wearlead malfunctionpacemakerpacemaker interrogation

Authors and institutions

  • Department of cardiology, The First Affiliated Hospital of Dalian Medical University, Dalian, China.

ORCID profiles: Peipei Ma, Lianjun Gao

Cite

APA

Dai, B., Yin, W., Ma, C., Ma, P., Wang, Y., Hong, L., Yu, X., Yin, X., Dong, Y., Xia, Y., Yang, Y., & Gao, L. (2024). Beware of atrial pacing‐induced ventricular depolarization: A case of lead malfunction. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14993

Vancouver

Dai B, Yin W, Ma C, Ma P, Wang Y, Hong L, et al. Beware of atrial pacing‐induced ventricular depolarization: A case of lead malfunction. Pacing and Clinical Electrophysiology. 2024. doi:10.1111/pace.14993

BibTeX
@article{Dai2024,
  author = {Dai, Bailing and Yin, Wenqing and Ma, Chengming and Ma, Peipei and Wang, Yingqi and Hong, Li and Yu, Xiaohong and Yin, Xiaomeng and Dong, Yingxue and Xia, Yunlong and Yang, Yanzong and Gao, Lianjun},
  title = {Beware of atrial pacing‐induced ventricular depolarization: A case of lead malfunction},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2024},
  doi = {10.1111/pace.14993}
}
Start for free →