His potential injury as the end point of screwing by a continuous recording technique in His bundle pacing: A case report

Jinyan Zhong iD, Longfu Jiang iD

Pacing and Clinical Electrophysiology · 2024

DOI: 10.1111/pace.14967 ↗

At a glance

Fallbericht in Pacing and Clinical Electrophysiology, 2024. Written by Jinyan Zhong and 1 co-authors. Institution: Department of Cardiology, Ningbo No. 2 hospital, Ningbo, Zhejiang, China. Major MEDLINE topics: Bundle of His, Cardiac Pacing, Artificial, Electrocardiography. Cited 1 times so far (Crossref); 5 references. Funded by: Ningbo Public Service Technology Foundation, Zhejiang Provincial Public Service and Application Research Foundation.

AbstraktSubjectsBibliograafilised andmedCiteRelated studies

Abstract (original language)

Background His bundle pacing (HBP) engaged electrical activation of both ventricles by stimulating the His‐Purkinje network, which could avoid marked ventricles dyssynchrony. The lead was given three to five clockwise rotations at the site with the His potential to anchor the interventricular septum. In 2018, the Multicenter His Bundle Pacing Collaborative Working Group recommended that the His bundle capture threshold should be lower than 2.5 V/1 ms in non‐pacing‐dependent patients, and pacing‐dependent patients should have a lower adjacent ventricular capture threshold as self‐backup. Therefore, to avoid safety issues such as loss of capture caused by increased threshold, we believe that more stringent criteria should be adopted in patients with atrioventricular block (AVB). In previous studies, the connection cable needed to be disconnected during the screwing. When the procedure was finished, the performer found that the patients with His bundle injury could obtain a lower threshold than those without His bundle injury. Although no studies of new bundle branch block (BBB) or AVB by the acute His bundle injury was reported. However, It is worrying that the damage of His bundle seems random during the procedure. How to balance avoiding severe injury with a lower capture threshold? At present, we report a case of light His injury and lower His capture threshold under continuous intracardiac electrocardiogram monitoring.

Subjects

MeSH terms (MEDLINE)

Bundle of HisCardiac Pacing, ArtificialElectrocardiographyHumansMaleAtrioventricular BlockAgedBundle-Branch Block

Author keywords

continuous recordinghis bundle pacinghis potential injuryjohn jiang's connecting cable

Authors and institutions

  • Department of Cardiology, Ningbo No. 2 hospital, Ningbo, Zhejiang, China.

ORCID profiles: Jinyan Zhong, Longfu Jiang

Funding

  • Ningbo Public Service Technology Foundation
  • Zhejiang Provincial Public Service and Application Research Foundation

Cite

APA

Zhong, J., & Jiang, L. (2024). His potential injury as the end point of screwing by a continuous recording technique in His bundle pacing: A case report. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14967

Vancouver

Zhong J, Jiang L. His potential injury as the end point of screwing by a continuous recording technique in His bundle pacing: A case report. Pacing and Clinical Electrophysiology. 2024. doi:10.1111/pace.14967

BibTeX
@article{Zhong2024,
  author = {Zhong, Jinyan and Jiang, Longfu},
  title = {His potential injury as the end point of screwing by a continuous recording technique in His bundle pacing: A case report},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2024},
  doi = {10.1111/pace.14967}
}
Start for free →