Acute pacing threshold elevation during simultaneous Micra leadless pacemaker implantation and AV node ablation: Clinical cases, computer model and practical recommendations

Hugo-Pierre Racine, Nicolas Dognin, Yanzhu Zhao, Benoit Plourde, Christian Steinberg, Teresa Whitman, François Philippon iD

Pacing and Clinical Electrophysiology · 2023

DOI: 10.1111/pace.14814 ↗

At a glance

Journal article in Pacing and Clinical Electrophysiology, 2023. Written by Hugo-Pierre Racine and 6 co-authors. Institution: Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval (IUCPQ-UL), Quebec, Canada. Major MEDLINE topics: Pacemaker, Artificial, Atrioventricular Node, Catheter Ablation, Atrial Fibrillation. Cited 3 times so far (Crossref); 26 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

In patients with refractory atrial fibrillation (AF), atrioventricular nodal (AVN) ablation and permanent pacemaker implantation is recommended. The Micra Transcatheter Pacing System (Micra) is a single chamber leadless pacemaker (LPM) and thus offers the possibility of AV node (AVN) ablation in the same procedure. Pacing threshold (PT) elevation after radiofrequency (RF) ablation is a potential complication.

Methods

We conducted a single center retrospective cohort study. Patients implanted with a Micra ( n = 84) and concomitant or delayed AVN ablation ( n = 12) from 2014 to 2022 were included. Two cases of acute Micra PT elevation immediately following RF AVN ablation required device retrieval and implantation of a new Micra. Procedural characteristics and electrophysiological parameters were analyzed, and a computer model was performed to determine factors responsible for acute PT elevations.

Results

A total of 84 patients were included. Mean age was 74 ± 10 and 48% were women. Twelve patients (14%) underwent AVN ablation. Two patients had acute PT elevation requiring device retrieval despite no direct contact of the ablation catheter with the Micra. Computer modeling shows that significant dissipated power due to electrical field coupling can occur at the tip or ring electrode if the catheter is not kept at a safe distance (≥35 mm) from the Micra when a maximum power of 100 W is delivered.

Conclusion

Concurrent AVN ablation and Micra implantation is safe in most patients. To prevent acute PT elevation, keeping a safe distance of ≥35 mm from the tip and ring electrodes of the Micra and using lower power output may prevent this complication.

Subjects

MeSH terms (MEDLINE)

Pacemaker, ArtificialAtrioventricular NodeCatheter AblationAtrial FibrillationHumansFemaleAgedRetrospective StudiesMaleComputer SimulationEquipment Design

Author keywords

AV node ablationleadless pacemakerpacing threshold elevation

Authors and institutions

  • Institut universitaire de cardiologie et de pneumologie de Québec - Université Laval (IUCPQ-UL), Quebec, Canada.
  • Medtronic Inc, Cardiac Rhythm Management, Minneapolis, USA.

ORCID profiles: François Philippon

Cite

APA

Racine, H. P., Dognin, N., Zhao, Y., Plourde, B., Steinberg, C., Whitman, T., & Philippon, F. (2023). Acute pacing threshold elevation during simultaneous Micra leadless pacemaker implantation and AV node ablation: Clinical cases, computer model and practical recommendations. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14814

Vancouver

Racine HP, Dognin N, Zhao Y, Plourde B, Steinberg C, Whitman T, et al. Acute pacing threshold elevation during simultaneous Micra leadless pacemaker implantation and AV node ablation: Clinical cases, computer model and practical recommendations. Pacing and Clinical Electrophysiology. 2023. doi:10.1111/pace.14814

BibTeX
@article{Racine2023,
  author = {Racine, Hugo-Pierre and Dognin, Nicolas and Zhao, Yanzhu and Plourde, Benoit and Steinberg, Christian and Whitman, Teresa and Philippon, François},
  title = {Acute pacing threshold elevation during simultaneous Micra leadless pacemaker implantation and AV node ablation: Clinical cases, computer model and practical recommendations},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2023},
  doi = {10.1111/pace.14814}
}
Start for free →