Total Hip Arthroplasty Stems and Femur Strain During Implant Insertion: A Biomechanical Study of Single Taper, Double Taper, and Collared Monolithic Stems
Orthopaedic Surgery · 2025
DOI: 10.1111/os.70110 ↗Kurz gefasst
Fachartikel in der Zeitschrift Orthopaedic Surgery, 2025. Verfasst von Barrett Bruno Torre und 6 weiteren Autor:innen. Beteiligte Einrichtung: University of Connecticut Health Center, Department of Orthopaedic Surgery, Farmington, Connecticut, USA. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Hip Prosthesis, Prosthesis Design, Femur. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: University of Connecticut.
Abstract (Originalsprache)
Objective
In total hip arthroplasty (THA), intraoperative periprosthetic femoral fracture (IOPFF) is a significant concern, often occurring during femoral canal instrumentation and stem implantation due to proximal femur strain. However, the impact of different stem designs on this strain remains unclear. This study conducted a biomechanical analysis comparing strain patterns on the proximal femur during the implantation of three stem types: single taper, double taper, and collared stems. This study aims to explore if there is a difference in strain patterns placed on the proximal femur during the implantation of three different stem types: single taper, double taper, and collared stems?
Methods
There were 24 cadaveric femurs randomly assigned to three groups based on implant geometry: Smith and Nephew anthology (single taper), synergy (double taper), and polar (collared) stems. Strain gauges were placed on the proximal femur to measure strain during implantation in both horizontal and vertical directions. Peak strain, the difference between maximum strain at final impaction and baseline, was recorded. The Kruskal–Wallis test compared peak strain between stem designs.
Results
At the medial proximal femur, collared stems produced compressive strain (−276) in the horizontal vector, differing significantly from the tensile strain generated by single (41, p = 0.009) and double taper stems (218.5, p = 0.003). No significant strain difference existed between single and double taper stems at the medial proximal femur. At the lateral proximal femur, double taper stem impaction resulted in compressive strain (−69), significantly differing from the tensile strain produced by single taper (221, p = 0.024) and collared stem impaction (462, p = 0.009). No strain differences were observed in other tested areas.
Conclusion
This study highlights distinct strain patterns at the medial and lateral proximal femur depending on the stem type. Collared stems induce compressive strain at the medial proximal femur, while double taper stems result in compressive strain at the lateral proximal femur. Understanding these differences may help reduce IOPFF risk in THA procedures. Based on these findings, the use of collard stems would be preferable over single or double taper stems in elderly patients with suboptimal bone quality, as it reduces strain on the medial cortex and improves immediate stability.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University of Connecticut Health Center, Department of Orthopaedic Surgery, Farmington, Connecticut, USA.
ORCID-Profile: Jordan Andre Bauer
Förderung
- University of Connecticut
Zitieren
APA
Torre, B. B., Bauer, J. A., Wellington, I., Sedghi, T., Neumann, D., Lindsay, A., & Solovyova, O. (2025). Total Hip Arthroplasty Stems and Femur Strain During Implant Insertion: A Biomechanical Study of Single Taper, Double Taper, and Collared Monolithic Stems. Orthopaedic Surgery. https://doi.org/10.1111/os.70110
Vancouver
Torre BB, Bauer JA, Wellington I, Sedghi T, Neumann D, Lindsay A, et al. Total Hip Arthroplasty Stems and Femur Strain During Implant Insertion: A Biomechanical Study of Single Taper, Double Taper, and Collared Monolithic Stems. Orthopaedic Surgery. 2025. doi:10.1111/os.70110
BibTeX
@article{Torre2025,
author = {Torre, Barrett Bruno and Bauer, Jordan Andre and Wellington, Ian and Sedghi, Tannaz and Neumann, Dillon and Lindsay, Adam and Solovyova, Olga},
title = {Total Hip Arthroplasty Stems and Femur Strain During Implant Insertion: A Biomechanical Study of Single Taper, Double Taper, and Collared Monolithic Stems},
journal = {Orthopaedic Surgery},
year = {2025},
doi = {10.1111/os.70110}
}
EUVIMED