EUVIMEDEUVIMED entdecken
Open Access

Fractures Around Collared and Collarless Hip Stems Initiate in High Strain Regions: A Combined Cadaveric Experiment and Finite Element Analysis

Ryan T Helbock iD, Clarisse Zigan iD, Andrew Hughes iD, Joseph D Lipman iD, Timothy M Wright iD, Peter K Sculco iD, Elizabeth B Gausden iD, Sony Manandhar iD, Ghislain Maquer iD, Jeff Bischoff iD, Fernando J Quevedo Gonzalez iD

Journal of Orthopaedic Research · 2026

DOI: 10.1002/jor.70267 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Orthopaedic Research, 2026. Verfasst von Ryan T Helbock und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomechanics, Hospital for Special Surgery, New York City, New York, USA. Hauptthemen laut MEDLINE: Hip Prosthesis, Arthroplasty, Replacement, Hip, Periprosthetic Fractures, Femoral Fractures. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Zimmer Biomet.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Periprosthetic femoral fracture (PFF) is a common early complication after primary total hip arthroplasty. Collared stems reduce but do not eliminate the incidence of PFF, and exhibit variable biomechanical effectiveness, emphasizing our limited understanding of the local bone–implant interaction mechanics. Our goal was to elucidate the relationship between the local strains at the bone–implant interface and the experimental fracture patterns and loads for collared and collarless stems. Six pairs of women cadaveric femurs implanted with a collarless or collared stem were loaded to failure under simulated stumbling to determine the location of PFF and the load to fracture, which we related to the collar‐calcar separation. Corresponding specimen‐specific FE models were developed to determine the strain at the bone‐implant interface and to predict the fracture onset load and location of fracture, which we related to the experimental location of PFF and the load‐to‐fracture. Load to fracture was greater for collared stems and was inversely correlated with the collar‐calcar separation at the time of implantation ( r = −0.80, p = 0.055). Fractures occurred in areas of high strain. The experimental fracture location coincided with the first yielding cortical element in five cases. The load at which the first cortical element yielded (i.e., onset of fracture was moderately correlated with the experimental fracture load ( R 2 = 0.43, RMSE = 1231 N). Our results emphasize the importance of initial calcar contact to realize the benefits of the collar. Localized load transfer was a precursor of macroscopic fracture, which initiated in areas of high tensile strain.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hip ProsthesisArthroplasty, Replacement, HipPeriprosthetic FracturesFemoral FracturesHumansFemaleFinite Element AnalysisAgedCadaverBiomechanical PhenomenaStress, MechanicalAged, 80 and overProsthesis DesignMiddle Aged

Stichwörter der Autor:innen

bone–implant interfacecollared femoral stemsfinite element modelingfracture initiationperiprosthetic femoral fracturestrain‐based failure prediction

Autor:innen und Institutionen

  • Department of Biomechanics, Hospital for Special Surgery, New York City, New York, USA.
  • Adult Reconstruction and Joint Replacement Service, Hospital for Special Surgery, New York City, New York, USA.
  • Zimmer Biomet, Warsaw, Indiana, USA.

ORCID-Profile: Ryan T Helbock, Clarisse Zigan, Andrew Hughes, Joseph D Lipman, Timothy M Wright, Peter K Sculco, Elizabeth B Gausden, Sony Manandhar, Ghislain Maquer, Jeff Bischoff, Fernando J Quevedo Gonzalez

Förderung

  • Zimmer Biomet

Zitieren

APA

Helbock, R. T., Zigan, C., Hughes, A., Lipman, J. D., Wright, T. M., Sculco, P. K., Gausden, E. B., Manandhar, S., Maquer, G., Bischoff, J., & Quevedo Gonzalez, F. J. (2026). Fractures Around Collared and Collarless Hip Stems Initiate in High Strain Regions: A Combined Cadaveric Experiment and Finite Element Analysis. Journal of Orthopaedic Research. https://doi.org/10.1002/jor.70267

Vancouver

Helbock RT, Zigan C, Hughes A, Lipman JD, Wright TM, Sculco PK, et al. Fractures Around Collared and Collarless Hip Stems Initiate in High Strain Regions: A Combined Cadaveric Experiment and Finite Element Analysis. Journal of Orthopaedic Research. 2026. doi:10.1002/jor.70267

BibTeX
@article{Helbock2026,
  author = {Helbock, Ryan T and Zigan, Clarisse and Hughes, Andrew and Lipman, Joseph D and Wright, Timothy M and Sculco, Peter K and Gausden, Elizabeth B and Manandhar, Sony and Maquer, Ghislain and Bischoff, Jeff and Quevedo Gonzalez, Fernando J},
  title = {Fractures Around Collared and Collarless Hip Stems Initiate in High Strain Regions: A Combined Cadaveric Experiment and Finite Element Analysis},
  journal = {Journal of Orthopaedic Research},
  year = {2026},
  doi = {10.1002/jor.70267}
}
Kostenlos starten →