Bonding positional accuracy of attachments and marginal adaptation of in‐house aligners – A quality improvement laboratory study
Orthodontics & Craniofacial Research · 2024
DOI: 10.1111/ocr.12843 ↗Kurz gefasst
Fachartikel in der Zeitschrift Orthodontics & Craniofacial Research, 2024. Verfasst von Nathália Moraes Carvalho Barreto Brandão und 5 weiteren Autor:innen. Beteiligte Einrichtung: Pontifical Catholic University of Minas Gerais, Belo Horizonte, Minas Gerais, Brazil. Hauptthemen laut MEDLINE: Dental Marginal Adaptation, Polyethylene Glycols, Polyethylene Terephthalates. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 25 Literaturangaben. Gefördert von: Coordenação de Aperfeiçoamento de Pessoal de Nível Superior.
Abstract (Originalsprache)
Objectives
To evaluate the 3D accuracy of attachment positioning and the adaptation of aligners to attachments using in‐house templates made with either polyethylene terephthalate glycol (PETG) or ethylene‐vinyl acetate (EVA) and either pressure or vacuum thermoforming machines. Materials and Methods Overall, 140 test specimens were resin‐printed. Templates for the attachment bonding were made with 1‐mm EVA or 0.5‐mm PETG laminates. Orthodontic aligners were manufactured with 0.75‐mm PETG. The thermoplastification process was carried out using either vacuum or pressure machines. The positional differences between the virtual and bonded attachments were assessed in the X , Y and Z coordinates. The marginal adaptation between the aligners and the attachments was measured.
Results
Minor inaccuracies in the positioning of the attachments were observed in all combinations of thermoforming machines and plastic laminates used to fabricate the templates, mainly in the superior–inferior ( Z ) dimension. PETG performed better than EVA in the anterior region ( p < .05). No association was found between thermoplastification machines and the accuracy of the positioning of the attachments ( p > .05). While small misadaptations between the aligners and the attachments were observed, the EVA templates performed better than the PETG templates.
Conclusions
The inaccuracy of the attachment positioning and the misadaptation of the aligners to the attachments were slight. The vacuum and pressure thermoplastification machines showed no difference in attachment positioning accuracy. The PETG template was better than the EVA template in the anterior region, but the EVA attachments presented a better adaptation to the aligners than the PETG attachments.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Pontifical Catholic University of Minas Gerais, Belo Horizonte, Minas Gerais, Brazil.
- Department of Dentistry, Pontifical Catholic University of Minas Gerais, Belo Horizonte, Minas Gerais, Brazil.
ORCID-Profile: Raiane Machado Maia, Bernardo Quiroga Souki
Förderung
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
Zitieren
APA
Brandão, N. M. C. B., Maia, R. M., Gomes, V. M., Resende, C., Antunes, A. N. D. G., & Souki, B. Q. (2024). Bonding positional accuracy of attachments and marginal adaptation of in‐house aligners – A quality improvement laboratory study. Orthodontics & Craniofacial Research. https://doi.org/10.1111/ocr.12843
Vancouver
Brandão NMCB, Maia RM, Gomes VM, Resende C, Antunes ANDG, Souki BQ. Bonding positional accuracy of attachments and marginal adaptation of in‐house aligners – A quality improvement laboratory study. Orthodontics & Craniofacial Research. 2024. doi:10.1111/ocr.12843
BibTeX
@article{Brando2024,
author = {Brandão, Nathália Moraes Carvalho Barreto and Maia, Raiane Machado and Gomes, Victor de Morais and Resende, Carolina and Antunes, Alberto Nogueira da Gama and Souki, Bernardo Quiroga},
title = {Bonding positional accuracy of attachments and marginal adaptation of in‐house aligners – A quality improvement laboratory study},
journal = {Orthodontics & Craniofacial Research},
year = {2024},
doi = {10.1111/ocr.12843}
}
EUVIMED