Nasopharyngeal airway assistance improves esophageal intubation rates of high‐resolution esophageal manometry catheters
Neurogastroenterology & Motility · 2024
DOI: 10.1111/nmo.14824 ↗At a glance
Journal article in Neurogastroenterology & Motility, 2024. Written by Andrew Leopold and 2 co-authors. Institution: Division of Gastroenterology & Hepatology, Department of Medicine, University of Maryland School of Medicine, Baltimore, Maryland, USA. Major MEDLINE topics: Manometry, Esophageal Motility Disorders. Cited 6 times so far (Crossref); 18 references. The full text is freely available in PubMed Central. Funded by: NIDDK NIH HHS, National Institutes of Health.
Abstract (original language)
Background
High‐resolution esophageal manometry (HREM) is the gold standard test for esophageal motility disorders. Nasopharyngeal airway‐assisted insertion of the HREM catheter is a suggested salvage technique for failure from the inability to pass the catheter through the upper esophageal sphincter (UES). It has not been demonstrated that the nasopharyngeal airway improves procedural success rate.
Methods
Patients undergoing HREM between March 2019 and March 2023 were evaluated. Chart review was conducted for patient factors and procedural success rates before and after use of nasopharyngeal airway. Patients from March 2019 to May 2021 did not have nasopharyngeal airway available and were compared to patients from May 2021 to March 2023 who had the nasopharyngeal airway available. Key Results In total, 523 HREM studies were conducted; 234 occurred prior to nasopharyngeal airway availability, and 289 occurred with nasopharyngeal airway availability. There was no difference in HREM catheter UES intubation rates between periods when a nasopharyngeal airway attempt was considered procedural failure (85% vs. 85%, p = 0.9). Nasopharyngeal airway use after UES intubation failure lead to improved UES intubation rates (94% vs. 85%, p < 0.01). Thirty‐six patients that failed HREM catheter UES intubation had the procedure reattempted with a nasopharyngeal airway, 30 (83%) of which were successful. The nasopharyngeal airway assisted catheter UES intubation for failures attributed to nasal pain and hypersensitivity, gagging, coughing, and pharyngeal coiling. Conclusions & Inferences Utilization of the nasopharyngeal airway increased rates of UES intubation. When HREM catheter placement through the UES fails, placement of a nasopharyngeal airway can be trialed to overcome patient procedural intolerance.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Division of Gastroenterology & Hepatology, Department of Medicine, University of Maryland School of Medicine, Baltimore, Maryland, USA.
- Department of Internal Medicine, University of Maryland School of Medicine, Baltimore, Maryland, USA.
- Division of Gastroenterology & Hepatology, VA Maryland Healthcare System, Baltimore, Maryland, USA.
ORCID profiles: Andrew Leopold, Guofeng Xie
Funding
- NIDDK NIH HHS
- National Institutes of Health
Cite
APA
Leopold, A., Wu, A., & Xie, G. (2024). Nasopharyngeal airway assistance improves esophageal intubation rates of high‐resolution esophageal manometry catheters. Neurogastroenterology & Motility. https://doi.org/10.1111/nmo.14824
Vancouver
Leopold A, Wu A, Xie G. Nasopharyngeal airway assistance improves esophageal intubation rates of high‐resolution esophageal manometry catheters. Neurogastroenterology & Motility. 2024. doi:10.1111/nmo.14824
BibTeX
@article{Leopold2024,
author = {Leopold, Andrew and Wu, Angela and Xie, Guofeng},
title = {Nasopharyngeal airway assistance improves esophageal intubation rates of high‐resolution esophageal manometry catheters},
journal = {Neurogastroenterology & Motility},
year = {2024},
doi = {10.1111/nmo.14824}
}
EUVIMED