The utility of balloon‐based and standard impedance technology to diagnose esophageal mucosal diseases in children
Neurogastroenterology & Motility · 2024
DOI: 10.1111/nmo.14726 ↗At a glance
Journal article in Neurogastroenterology & Motility, 2024. Written by Rachel Rosen and 6 co-authors. Institution: Division of Gastroenterology, Hepatology and Nutrition, Aerodigestive Center, Boston Children's Hospital, Boston, Massachusetts, USA. Major MEDLINE topics: Gastroesophageal Reflux, Eosinophilic Esophagitis. Cited 1 times so far (Crossref); 9 references. The full text is freely available in PubMed Central. Funded by: NIDDK NIH HHS, NIH HHS, National Institutes of Health.
Topics: Kinderkrankenpflege
Abstract (original language)
Background
Standard impedance catheters and balloon‐based mucosal impedance catheters (BBMI) have been used to assess mucosal integrity and diagnose mucosal diseases. The goal of this study was to determine the age‐related technical issues associated with mucosal balloon inflation, validate the BBMI measurement against a standard impedance probe, and compare software‐generated diagnoses to histologic diagnoses.
Methods
We prospectively recruited patients undergoing endoscopy, during which patients underwent standard mucosal impedance catheters and BBMI measurements. Measurements were compared to each other, to the histologic diagnoses, and to the number of eosinophils per high power field. We then compared the patients' diagnosis to that assigned by the BBMI software. Key Results Sixty‐two patients (mean age: 62 ± 62 months) were recruited, including non‐GERD ( N = 40), GERD ( N = 15), and EoE ( N = 7) patients. There were significant differences between the impedance values measured by the two technologies at each esophageal height ( p < 0.003). There were significant correlations between the mean impedance values taken by the two catheters in the distal ( r 2 = 0.272, p = 0.04), mid ( r 2 = 0.371, p < 0.001), and proximal ( r 2 = 0.259, p = 0.05) esophagus. There were significant differences in BBMI impedance values across diagnoses in the mid and proximal esophagus ( p = 0.024 and 0.025, respectively). While not statistically significant ( p = 0.061–0.073), the standard catheter showed similar trends by diagnosis. Using the BBMI diagnostic prediction software, 33%–72% of patients were misclassified. Conclusion and Inferences While there was significant variability in impedance values between technologies within patients, regional measurements were consistent across catheters. Automated analyses lacked the sensitivity to diagnose inflammatory disorders.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Division of Gastroenterology, Hepatology and Nutrition, Aerodigestive Center, Boston Children's Hospital, Boston, Massachusetts, USA.
- Institutional Centers for Clinical and Translational Research, Boston Children's Hospital, Boston, Massachusetts, USA.
ORCID profiles: Rachel Rosen, Samuel Nurko
Funding
- NIDDK NIH HHS
- NIH HHS
- National Institutes of Health
Cite
APA
Rosen, R., Du, M., Akkara, A., Catacora, A., Wen, A., Liu, E., & Nurko, S. (2024). The utility of balloon‐based and standard impedance technology to diagnose esophageal mucosal diseases in children. Neurogastroenterology & Motility. https://doi.org/10.1111/nmo.14726
Vancouver
Rosen R, Du M, Akkara A, Catacora A, Wen A, Liu E, et al. The utility of balloon‐based and standard impedance technology to diagnose esophageal mucosal diseases in children. Neurogastroenterology & Motility. 2024. doi:10.1111/nmo.14726
BibTeX
@article{Rosen2024,
author = {Rosen, Rachel and Du, Maritha and Akkara, Anna and Catacora, Andrea and Wen, Anna and Liu, Enju and Nurko, Samuel},
title = {The utility of balloon‐based and standard impedance technology to diagnose esophageal mucosal diseases in children},
journal = {Neurogastroenterology & Motility},
year = {2024},
doi = {10.1111/nmo.14726}
}
EUVIMED