Development of an Imaging‐Based Method for Analyzing Voided Urine Flow Using Conventional and High‐Speed Video Cameras: A Phantom Study
LUTS: Lower Urinary Tract Symptoms · 2026
DOI: 10.1111/luts.70084 ↗Kurz gefasst
Fachartikel in der Zeitschrift LUTS: Lower Urinary Tract Symptoms, 2026. Verfasst von Masahiro Goto und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, Shinshu University School of Medicine, Matsumoto, Japan. Hauptthemen laut MEDLINE: Urodynamics, Video Recording, Urethra, Urination. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Japan Society for the Promotion of Science.
Abstract (Originalsprache)
Aims
This study examined whether voided urine form could be an independent parameter to evaluate pathology of the lower urinary tract separately from conventional uroflowmetric measurements. We also evaluated a novel combined methodology of dynamically measuring the velocity and cross‐sectional area (CA) of voided urine.
Methods
A male urethral phantom was connected to an electric pump, and voided water was recorded using a high‐speed video camera (HSVC). The parameters of voided water form included length to the first twist (LFT), length to the scattered water (LSW), and voided water angle (VWA), which were measured in combination with conventional uroflowmetry. The velocity and CA of voided water were measured using an analysis program of fluid movement. Variations of ductal deformity were simulated using silicon tube attachments.
Results
Each parameter increased with pump speed apart from VWA. Flow rate did not change for any attachment except for the stricture type, for which the relative change of LFT surpassed that of LSW. The VWA results were slightly increased by urethral deformity. All results for velocity and CA increased with pump speed and were unaffected by urethral deformity under fixed water flow.
Conclusion
LFT and LSW represent novel voided water form parameters that are separate from conventional uroflowmetric measurements, such as flow rate. Voided water velocity and CA can be non‐invasively determined by the combination of uroflowmetry and a HSVC.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Urology, Shinshu University School of Medicine, Matsumoto, Japan.
ORCID-Profile: Masahiro Goto, Tomonori Minagawa, Tetsuichi Saito, Tetsuya Imamura, Yoshiyuki Akiyama
Förderung
- Japan Society for the Promotion of Science
Zitieren
APA
Goto, M., Minagawa, T., Hara, H., Ogawa, N., Saito, T., Imamura, T., Ueno, M., & Akiyama, Y. (2026). Development of an Imaging‐Based Method for Analyzing Voided Urine Flow Using Conventional and High‐Speed Video Cameras: A Phantom Study. LUTS: Lower Urinary Tract Symptoms. https://doi.org/10.1111/luts.70084
Vancouver
Goto M, Minagawa T, Hara H, Ogawa N, Saito T, Imamura T, et al. Development of an Imaging‐Based Method for Analyzing Voided Urine Flow Using Conventional and High‐Speed Video Cameras: A Phantom Study. LUTS: Lower Urinary Tract Symptoms. 2026. doi:10.1111/luts.70084
BibTeX
@article{Goto2026,
author = {Goto, Masahiro and Minagawa, Tomonori and Hara, Hiroaki and Ogawa, Noriyuki and Saito, Tetsuichi and Imamura, Tetsuya and Ueno, Manabu and Akiyama, Yoshiyuki},
title = {Development of an Imaging‐Based Method for Analyzing Voided Urine Flow Using Conventional and High‐Speed Video Cameras: A Phantom Study},
journal = {LUTS: Lower Urinary Tract Symptoms},
year = {2026},
doi = {10.1111/luts.70084}
}
EUVIMED