Influence of transurethral catheters on urine pressure-flow relationships in males: A computational fluid-dynamics study
Computer Methods and Programs in Biomedicine · 2023
DOI: 10.1016/j.cmpb.2023.107594 ↗Kurz gefasst
Fachartikel in der Zeitschrift Computer Methods and Programs in Biomedicine, 2023. Verfasst von Maria Vittoria Mascolini und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Industrial Engineering, University of Padova, Padova, Italy; Centre of Mechanics of Biological Materials, University of Padova, Padova,…. Hauptthemen laut MEDLINE: Urinary Bladder, Urethra. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministero dell’Istruzione, dell’Università e della Ricerca, Università degli Studi di Padova.
Abstract (Originalsprache)
BACKGROUND AND OBJECTIVE: In the field of urology, the pressure-flow study (PFS) is an essential urodynamics practise which requires the patient's transurethral catheterization during the voiding phase of micturition to evaluate the functionality of the lower urinary tract (LUT) and reveal the pathophysiology of its dysfunctionality. However, the literature evidences confusion regarding the interference of the catheterization on the urethral pressure-flow behaviour. METHODS: The present research study represents the first Computational Fluid-Dynamics (CFD) approach to this urodynamics issue, analysing the influence of a catheter in the male LUT through case studies which included the inter-individual and intra-individual dependence. A set of four three dimensional (3D) models of the male LUT, different in urethral diameters, and a set of three 3D models of the transurethral catheter, diverse in calibre, were developed leading to 16 CFD non-catheterized either catheterized configurations, to describe the typical micturition scenario considering both urethra and catheter characteristics. RESULTS: The developed CFD simulations showed that the urine flow field during micturition was influenced by the urethral cross-sectional area and each catheter determined a specific decrease in flow rate if compared to the relative free uroflow. CONCLUSIONS: In-silico methods allow to analyse relevant urodynamics aspects, which could not be investigated in vivo, and may support the clinical PFS to reduce uncertainty on urodynamic diagnosis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Industrial Engineering, University of Padova, Padova, Italy; Centre of Mechanics of Biological Materials, University of Padova, Padova, Italy.
- Centre of Mechanics of Biological Materials, University of Padova, Padova, Italy; Department of Civil, Environmental and Architectural Engineering, University of Padova, Padova, Italy; Department of Biomedical Sciences,…
Förderung
- Ministero dell’Istruzione, dell’Università e della Ricerca
- Università degli Studi di Padova
Zitieren
APA
Mascolini, M. V., Fontanella, C. G., Berardo, A., & Carniel, E. L. (2023). Influence of transurethral catheters on urine pressure-flow relationships in males: A computational fluid-dynamics study. Computer Methods and Programs in Biomedicine. https://doi.org/10.1016/j.cmpb.2023.107594
Vancouver
Mascolini MV, Fontanella CG, Berardo A, Carniel EL. Influence of transurethral catheters on urine pressure-flow relationships in males: A computational fluid-dynamics study. Computer Methods and Programs in Biomedicine. 2023. doi:10.1016/j.cmpb.2023.107594
BibTeX
@article{Mascolini2023,
author = {Mascolini, Maria Vittoria and Fontanella, Chiara Giulia and Berardo, Alice and Carniel, Emanuele Luigi},
title = {Influence of transurethral catheters on urine pressure-flow relationships in males: A computational fluid-dynamics study},
journal = {Computer Methods and Programs in Biomedicine},
year = {2023},
doi = {10.1016/j.cmpb.2023.107594}
}