The Accuracy of a Commercial Wearable Neonatal Home‐Monitoring Device: A Simulation Study
Journal of Paediatrics and Child Health · 2026-08 · 62 (8) · 1509-1519
DOI: 10.1111/jpc.70471 ↗At a glance
Journal article in Journal of Paediatrics and Child Health, published online on 2026-06-15. Written by Amelia N Noone and 5 co-authors. Institution: Robinson Research Institute, College of Health, Adelaide University, Adelaide, South Australia, Australia. Major MEDLINE topics: Wearable Electronic Devices, Accelerometry, Apnea. 15 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Objective
Wearable newborn‐monitoring devices are increasingly popular and widely available for home use. One approach includes the use of accelerometer detection of breathing movements. We investigated the accuracy of breathing and apnoea detection by a wearable accelerometer in a simulated environment.
Design
This study used simulation as a method to evaluate the performance of a wearable accelerometer‐based breathing detection device. A term mannequin fitted with an accelerometer suit wirelessly linked to a mobile app (Goldilocks) was NeoPuff ventilated through a cuffed endotracheal tube. Peak inspiratory pressure (PIP) levels of 5–25 cmH 2 O were delivered with zero end expiratory pressure. Each trial consisted of 30 s of ventilation, followed by simulated apnoea ( n = 150). App‐based detection and alert times were recorded blinded to ventilation status. Detection accuracy across PIP levels was analyzed using Kruskal–Wallis tests followed by Dunn's multiple comparisons with Bonferroni correction.
Results
PIP level affected both apnoea detection (H(4) = 58.9, p < 0.0001) and breathing detection (H(4) = 86.5, p < 0.001) with performance reduced at 5 cmH 2 O compared to higher PIP. Apnoea and breathing detection were most accurate at 20 cmH 2 O. Mean (SD) apnoea detection alarm times ranged between 31(0) and 60(22.6) seconds, with no difference between PIP group (H(4) = 9.3, p > 0.05).
Conclusions
The accelerometer displayed pressure‐dependent accuracy, increasing with higher PIP, suggesting a minimum pressure threshold is needed for reliable monitoring. App alert times varied greatly, compounding the accuracy of adverse event detection. To determine their true utility, rigorous investigation of neonatal wearable home‐monitoring devices is required.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Robinson Research Institute, College of Health, Adelaide University, Adelaide, South Australia, Australia.
- Goldilocks Suit, Adelaide, South Australia, Australia.
- Department of Neonatal Medicine, Women's and Children's Hospital, Adelaide, South Australia, Australia.
ORCID profiles: Amelia N Noone, Michael J Stark
Cite
APA
Noone, A. N., Bailey, D. N., Crawford, T. M., Richards, S., Andersen, C. C., & Stark, M. J. (2026). The Accuracy of a Commercial Wearable Neonatal Home‐Monitoring Device: A Simulation Study. Journal of Paediatrics and Child Health, 62(8), 1509-1519. https://doi.org/10.1111/jpc.70471
Vancouver
Noone AN, Bailey DN, Crawford TM, Richards S, Andersen CC, Stark MJ. The Accuracy of a Commercial Wearable Neonatal Home‐Monitoring Device: A Simulation Study. Journal of Paediatrics and Child Health. 2026;62(8):1509-1519. doi:10.1111/jpc.70471
BibTeX
@article{Noone2026,
author = {Noone, Amelia N and Bailey, Danielle N and Crawford, Tara M and Richards, Shem and Andersen, Chad C and Stark, Michael J},
title = {The Accuracy of a Commercial Wearable Neonatal Home‐Monitoring Device: A Simulation Study},
journal = {Journal of Paediatrics and Child Health},
year = {2026},
volume = {62},
number = {8},
pages = {1509--1519},
doi = {10.1111/jpc.70471}
}
EUVIMED