Biosensor Compared With Standard Care to Identify Extravasation Injuries in Paediatric Intensive Care: A Randomised Controlled Trial
Journal of Paediatrics and Child Health · 2026-06 · 62 (6) · 1034-1043
DOI: 10.1111/jpc.70378 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Journal of Paediatrics and Child Health, online erschienen am 14.04.2026. Verfasst von Craig A McBride und 10 weiteren Autor:innen. Beteiligte Einrichtung: Surgical Team: Infants, Toddlers, Children (STITCh), Queensland Children's Hospital, South Brisbane, Queensland, Australia. Hauptthemen laut MEDLINE: Extravasation of Diagnostic and Therapeutic Materials, Biosensing Techniques, Catheterization, Peripheral. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Children's Hospital Foundation, ivWatch LLC.
Themen: Intensivpflege
Abstract (Originalsprache)
Aim
To evaluate whether, in children treated in paediatric intensive care units (PICU) with peripheral venous access devices (PVADs) receiving intermediate‐ or high‐risk peripheral intravenous medications, the use of a tissue‐infiltration biosensor (ivWatch), compared to standard care, reduces extravasation injuries.
Methods
A superiority, parallel group randomised controlled trial conducted in a single Queensland quaternary hospital PICU from September 2020 to July 2022. Children with PVADs inserted in the distal half of the limb and receiving intermediate‐ to high‐risk infusates for ≥ 24 h were randomly allocated in a 1:1 ratio to biosensor (ivWatch monitor applied within 10–15 mm of the PVAD tip) plus standard clinical observation, or standard observation alone. Primary outcome was extravasation severity on the 4‐point Cincinnati Children's Hospital Medical Centre Extravasation Harm Scale. Injuries were compared between treatment groups using ordinal logistic regression.
Results
A total of 175 children were enrolled, 88 to biosensor and 87 to standard care. Extravasation severity was similar between groups, with no, mild and moderate injury occurring in 56 (69%), 25 (30%) and 2 (2%) participants in the biosensor group and 61 (70%), 24 (28%) and 2 (2%) in the standard care group (proportional odds ratio = 1.13; 95% CI: 0.59, 2.15, p = 0.71). No statistically significant between‐group differences in extravasation severity, volume, treatment sequelae, PVADs used or dwell time were observed. The biosensor was acceptable to nurses.
Conclusions
In this PICU population, biosensor monitoring in addition to standard care did not significantly reduce extravasation injuries. The technology was acceptable to bedside nursing staff. Trial Registration: ACTRN12620000317998
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Surgical Team: Infants, Toddlers, Children (STITCh), Queensland Children's Hospital, South Brisbane, Queensland, Australia.
- NHMRC Centre for Wiser Wound Care, Griffith University, Nathan, Queensland, Australia.
- Pegg Leditschke Children's Burns Centre, Children's Health Queensland, South Brisbane, Queensland, Australia.
- Child Health Research Centre, the University of Queensland, South Brisbane, Queensland, Australia.
- Paediatric Intensive Care Unit, Queensland Children's Hospital, South Brisbane, Queensland, Australia.
- Department of Intensive Care Medicine and Neonatology, Children's Research Center, University Children's Hospital Zürich, University of Zurich, Zurich, Switzerland.
ORCID-Profile: Craig A McBride, Sarfaraz Rahiman, Luregn J Schlapbach, Jessica A Schults, Tricia M Kleidon, Rebecca Paterson, Joshua Byrnes, Robert S Ware, Amanda J Ullman
Förderung
- Children's Hospital Foundation
- ivWatch LLC
Zitieren
APA
McBride, C. A., Rahiman, S., Kennedy, M., Schlapbach, L. J., Schults, J. A., Kleidon, T. M., Baveas, T., Paterson, R., Byrnes, J., Ware, R. S., & Ullman, A. J. (2026). Biosensor Compared With Standard Care to Identify Extravasation Injuries in Paediatric Intensive Care: A Randomised Controlled Trial. Journal of Paediatrics and Child Health, 62(6), 1034-1043. https://doi.org/10.1111/jpc.70378
Vancouver
McBride CA, Rahiman S, Kennedy M, Schlapbach LJ, Schults JA, Kleidon TM, et al. Biosensor Compared With Standard Care to Identify Extravasation Injuries in Paediatric Intensive Care: A Randomised Controlled Trial. Journal of Paediatrics and Child Health. 2026;62(6):1034-1043. doi:10.1111/jpc.70378
BibTeX
@article{McBride2026,
author = {McBride, Craig A and Rahiman, Sarfaraz and Kennedy, Melanie and Schlapbach, Luregn J and Schults, Jessica A and Kleidon, Tricia M and Baveas, Thim and Paterson, Rebecca and Byrnes, Joshua and Ware, Robert S and Ullman, Amanda J},
title = {Biosensor Compared With Standard Care to Identify Extravasation Injuries in Paediatric Intensive Care: A Randomised Controlled Trial},
journal = {Journal of Paediatrics and Child Health},
year = {2026},
volume = {62},
number = {6},
pages = {1034--1043},
doi = {10.1111/jpc.70378}
}