A novel dental implant nut system designed to enhance primary stability: Mechanical and finite element analysis
Journal of Prosthodontics · 31.05.2026
DOI: 10.1111/jopr.70169 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Prosthodontics, online erschienen am 31.05.2026. Verfasst von Mert Şahin und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Erciyes University, Kayseri, Türkiye. Die Arbeit nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: The Scientific and Technological Research Council of Türkiye, TÜBİTAK.
Abstract (Originalsprache)
Purpose
The present study aimed to evaluate, in a preclinical setting, the biomechanical behavior and mechanical performance of a newly developed dental implant nut system designed to address primary stability challenges in the severely resorbed posterior maxilla. Materials and Methods The evaluation consisted of two complementary stages: an in vitro mechanical experiment and a finite element analysis (FEA). In the experimental phase, polyurethane jaw models simulating a severely atrophic posterior maxilla with approximately 1 mm residual bone height were used. Four groups were prepared: single implant without nut, single implant with nut, double implant without nut, and double implant with nut ( n = 10 per group). Primary stability was measured using resonance frequency analysis (RFA) with an Osstell device. In the computational phase, three‐dimensional finite element models were constructed from cone‐beam computed tomography (CBCT) data of the same configuration to evaluate detachment forces and stress distribution within the bone–implant complex under vertical loading.
Results
In single‐implant models, mean implant stability quotient (ISQ) values increased from 10.91 ± 7.35 to 18.27 ± 7.11 after nut application. In double‐implant models, ISQ values increased from 11.13 ± 4.14 to 19.24 ± 4.24 ( p < 0.05). FEA results revealed that the detachment force increased from 16.5 to 20.33 N in single‐implant models and from 15 to 22.3 N in double‐implant models. Compressive stresses on the bone were lower in nut‐supported configurations, indicating a more favorable load distribution.
Conclusion
The nut system was associated with increased implant primary stability and a more favorable stress distribution under the conditions of this preclinical study. These findings should be interpreted as preliminary biomechanical evidence, as no in vivo or clinical validation was performed. Therefore, this design may be considered a mechanical stabilization concept that warrants further investigation through in vivo and clinical studies.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Erciyes University, Kayseri, Türkiye.
- Dentbiochem Co., Erciyes Teknopark, Kayseri, Türkiye.
- Department of Mechanical Engineering, Faculty of Engineering, Abdullah Gül University, Kayseri, Türkiye.
- Private Practice, Istanbul, Türkiye.
ORCID-Profile: Mert Şahin
Förderung
- The Scientific and Technological Research Council of Türkiye
- TÜBİTAK
Zitieren
APA
Şahin, M., Soylu, E., Demirbaş, A. E., Bal, B., & Alkan, A. (2026). A novel dental implant nut system designed to enhance primary stability: Mechanical and finite element analysis. Journal of Prosthodontics. https://doi.org/10.1111/jopr.70169
Vancouver
Şahin M, Soylu E, Demirbaş AE, Bal B, Alkan A. A novel dental implant nut system designed to enhance primary stability: Mechanical and finite element analysis. Journal of Prosthodontics. 2026. doi:10.1111/jopr.70169
BibTeX
@article{ahin2026,
author = {Şahin, Mert and Soylu, Emrah and Demirbaş, Ahmet Emin and Bal, Burak and Alkan, Alper},
title = {A novel dental implant nut system designed to enhance primary stability: Mechanical and finite element analysis},
journal = {Journal of Prosthodontics},
year = {2026},
doi = {10.1111/jopr.70169}
}
EUVIMED