Fachartikel in der Zeitschrift Journal of Prosthodontics, 2025. Verfasst von Mario Ceddia und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanics, Mathematics and Management, Polytechnic University of Bari, Bari, Italy. Hauptthemen laut MEDLINE: Finite Element Analysis, Dental Stress Analysis, Dental Implants, Dental Prosthesis Design. Die Arbeit nennt 88 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
This study aimed to assess the response of peri‐implant tissues, both hard and soft, to mechanical stress when using a primary healing implant (PHI) with two different transmucosal profiles: concave (Model A) and divergent (Model B). The investigation also sought to observe bone modeling under post‐extraction conditions. Materials and Methods The methodology involved the creation of a three‐dimensional bone model of the first molar region, derived from a computed tomography scan. Subsequently, two implants were inserted into the bone site and subjected to a loading force of 100 N at a 45° angle.
Results
The results of stress analysis, using the von Mises criterion, indicated that Model A exhibited a more uniform stress distribution within the soft tissues, registering a maximum value of 75 MPa, in contrast to 126 MPa observed in Model B. Concerning implant stress, the peak value was recorded at the conometric connection zone between the implant and the abutment, measuring 138 MPa for Model B and 125 MPa for Model A. The study specifically analyzed cortical bone stress, which revealed levels of 72 MPa for Model B and 64 MPa for Model A. Additionally, stress distribution in immature bone ranged from 1.3 to 9 MPa for Model A and from 1.5 to 12 MPa for Model B.
Conclusions
The finite element method represents a valuable tool for the design and optimization of implant shapes, taking into account occlusal loads and specific anatomical locations. This approach aims to enhance the stimulation of both soft and hard tissues, thereby mitigating the risk of implant failure.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Mechanics, Mathematics and Management, Polytechnic University of Bari, Bari, Italy.
- Department of Medical, Oral and Biotechnological Sciences, "G. D'Annunzio" University of Chieti-Pescara, Chieti, Italy.
- Center for Advanced Studies and Technology (CAST), "G. D'Annunzio" University of Chieti-Pescara, Chieti, Italy.
- Division of Dental Research Administration, Tufts University School of Dental Medicine, Boston, Massachusetts, USA.
- Independent Researcher, San Vendemiano, Italy.
- Independent Researcher, Sciacca, Italy.
ORCID-Profile: Tea Romasco, Natalia Di Pietro
Zitieren
APA
Ceddia, M., Romasco, T., Comuzzi, L., Cipollina, A., Piattelli, A., Di Pietro, N., & Trentadue, B. (2025). Finite element analysis of primary healing implants with different transmucosal designs. Journal of Prosthodontics. https://doi.org/10.1111/jopr.14044
Vancouver
Ceddia M, Romasco T, Comuzzi L, Cipollina A, Piattelli A, Di Pietro N, et al. Finite element analysis of primary healing implants with different transmucosal designs. Journal of Prosthodontics. 2025. doi:10.1111/jopr.14044
BibTeX
@article{Ceddia2025,
author = {Ceddia, Mario and Romasco, Tea and Comuzzi, Luca and Cipollina, Alessandro and Piattelli, Adriano and Di Pietro, Natalia and Trentadue, Bartolomeo},
title = {Finite element analysis of primary healing implants with different transmucosal designs},
journal = {Journal of Prosthodontics},
year = {2025},
doi = {10.1111/jopr.14044}
}
EUVIMED