EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY-NC 4.0

Accuracy of digital duplication scanning methods for complete dentures

Abdullah Alehaideb, Wei-Shao Lin iD, John A Levon, Tien-Min G Chu iD, Chao-Chieh Yang iD

Journal of Prosthodontics · 2025-01 · 34 (1) · 42-48

DOI: 10.1111/jopr.13788 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Prosthodontics, online erschienen am 27.11.2023. Verfasst von Abdullah Alehaideb und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics, Indiana University School of Dentistry, Indianapolis, Indiana, USA. Hauptthemen laut MEDLINE: Denture, Complete, Denture Design. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To compare the accuracy of four digital scanning methods in duplicating a complete denture. Material and Methods Four scanning methods were used: cone beam computed tomography (CBCT), Straumann desktop scanner (DS), Trios intraoral scanner (TIO), and Virtuo Vivo intraoral scanner (VVIO). Each method was used to duplicate all the surfaces of a printed complete denture. The denture was scanned 10 times in each group. The trueness (in root mean square, RMS) and precision (in standard deviation, SD) were calculated by comparing the combined dentition, denture extension, and intaglio surfaces with the reference file. One‐way analysis of variance and F ‐tests were used to test statistical differences ( α = 0.05).

Results

For the scanning accuracy of the whole denture, CBCT showed the highest RMS (0.249 ± 0.020 mm) and lowest trueness than DS (0.124 ± 0.014 mm p < 0.001), TIO (0.131 ± 0.006 mm p < 0.001), and VVIO (0.227 ± 0.020 mm p = 0.017), while DS and TIO showed smaller RMS than VVIO ( p < 0.001). For the trueness of dentition, denture extension, and intaglio surfaces, CBCT also showed the highest mean RMS and lowest trueness among all groups ( p < 0.001). DS and TIO had smaller mean RMS and higher trueness among all groups in all surfaces ( p < 0.001, except VVIO in intaglio surface, p > 0.05). TIO had significantly lower within‐group variability of RMS and highest precision compared to DS ( p = 0.013), CBCT ( p = 0.001), and VVIO ( p < 0.001) in the combined surface. For dentition and denture extension surfaces, TIO showed similar within‐group variability of RMS with the DS group ( p > 0.05) and lower than CBCT and VVIO ( p < 0.001).

Conclusion

The 7 Series desktop scanner and Trios 4 intraoral scanner can duplicate dentures in higher trueness than CBCT and the Virtuo Vivo intraoral scanner. The Trios 4 intraoral scanner was more precise in the combined surfaces than other scanning methods, while the 7 Series desktop scanner and Trios 4 intraoral scanner were more precise in the denture extension surface.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Denture, CompleteDenture DesignCone-Beam Computed TomographyHumansComputer-Aided Design

Stichwörter der Autor:innen

denture duplicationdigital dentistryintraoral scanprecisiontrueness

Autor:innen und Institutionen

  • Department of Prosthodontics, Indiana University School of Dentistry, Indianapolis, Indiana, USA.
  • Department of Biomedical Sciences and Comprehensive Care, Indiana University School of Dentistry, Indianapolis, Indiana, USA.

ORCID-Profile: Wei-Shao Lin, Tien-Min G Chu, Chao-Chieh Yang

Zitieren

APA

Alehaideb, A., Lin, W. S., Levon, J. A., Chu, T. G., & Yang, C. C. (2025). Accuracy of digital duplication scanning methods for complete dentures. Journal of Prosthodontics, 34(1), 42-48. https://doi.org/10.1111/jopr.13788

Vancouver

Alehaideb A, Lin WS, Levon JA, Chu TG, Yang CC. Accuracy of digital duplication scanning methods for complete dentures. Journal of Prosthodontics. 2025;34(1):42-48. doi:10.1111/jopr.13788

BibTeX
@article{Alehaideb2025,
  author = {Alehaideb, Abdullah and Lin, Wei-Shao and Levon, John A and Chu, Tien-Min G and Yang, Chao-Chieh},
  title = {Accuracy of digital duplication scanning methods for complete dentures},
  journal = {Journal of Prosthodontics},
  year = {2025},
  volume = {34},
  number = {1},
  pages = {42--48},
  doi = {10.1111/jopr.13788}
}
Kostenlos starten →