Influence of scan technology on the accuracy and speed of intraoral scanning systems for the edentulous maxilla: An in vitro study
Journal of Prosthodontics · 2023
DOI: 10.1111/jopr.13633 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Prosthodontics, 2023. Verfasst von Reham Bassuni Osman und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics, Faculty of Dentistry, Cairo University, Cairo, Egypt. Hauptthemen laut MEDLINE: Imaging, Three-Dimensional, Mouth, Edentulous. Die Arbeit wurde bisher 35-mal zitiert (Crossref) und nennt 34 Literaturangaben.
Abstract (Originalsprache)
Purpose
To compare the accuracy of three intraoral scanners in terms of trueness and precision relative to the scanner acquisition technology and scan capture mode. Scan speed of each scanner was also evaluated. Materials and methods An edentulous maxillary arch was digitized (reference model) and 3D‐printed using an SLA‐based 3D‐printer (XFAB; DWS, Italy) ( n = 10). Each model was scanned using three intraoral scanners, each with different scan technologies: confocal (Trios 3; 3Shape, Copenhagen, Denmark), parallel confocal (iTero; Align Technology), and triangulation (Medit i700). Scan time and scanning accuracy (trueness and precision) were calculated using digital subtraction technique (Geomagic Control X v2020, 3DSystems, USA). One‐way analysis of variance (ANOVA) test was used to detect differences in trueness, precision, and scanning time between the tested groups ( p < 0.05).
Results
ANOVA results showed statistically significant differences in trueness, precision, and scan time among the tested groups. Confocal scanning technique (Trios 3) showed the highest trueness and precision (root mean square estimate [RMSE] 0.094 and 0.096, respectively) followed by iTero displaying parallel confocal technique (RMSE 0.113 and 0.133, respectively); the difference was not significant ( p = 0.849, p = 0.488). Further, Trios showed the longest scanning time (100 s) compared to iTero and Medit i700 ( p = 0.011 and 0.002, respectively). Medit i700 presenting triangulation scan technology revealed lowest trueness and precision (RMSE 0.268) ( p = 0.000, p = 0.001) and fastest scan time (59 s) close to iTero (66 s) ( p = 0.802).
Conclusions
Scanner technology had an influence on the accuracy and scan speed of the acquired intraoral scans. The Trios 3 scanner featuring the confocal acquisition technology displayed the highest trueness, precision, and longest scan time. Medit i700 IOS with triangulation acquisition concept featured the lowest trueness, precision, and fastest scan speed. There is no ideal scanner with the best combination of accuracy and scan speed.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Prosthodontics, Faculty of Dentistry, Cairo University, Cairo, Egypt.
- Department of Prosthetic Dental Sciences, College of Dentistry, King Saud University, Riyadh, Saudi Arabia.
ORCID-Profile: Reham Bassuni Osman, Nawal Murshed Alharbi
Zitieren
APA
Osman, R. B., & Alharbi, N. M. (2023). Influence of scan technology on the accuracy and speed of intraoral scanning systems for the edentulous maxilla: An in vitro study. Journal of Prosthodontics. https://doi.org/10.1111/jopr.13633
Vancouver
Osman RB, Alharbi NM. Influence of scan technology on the accuracy and speed of intraoral scanning systems for the edentulous maxilla: An in vitro study. Journal of Prosthodontics. 2023. doi:10.1111/jopr.13633
BibTeX
@article{Osman2023,
author = {Osman, Reham Bassuni and Alharbi, Nawal Murshed},
title = {Influence of scan technology on the accuracy and speed of intraoral scanning systems for the edentulous maxilla: An in vitro study},
journal = {Journal of Prosthodontics},
year = {2023},
doi = {10.1111/jopr.13633}
}
EUVIMED