Pitfalls in 2HG Detection With TE‐Optimized MRS at 3T
Journal of Neuroimaging · 2026
DOI: 10.1111/jon.70146 ↗At a glance
Journal article in Journal of Neuroimaging, 2026. Written by Seyma Alcicek and 13 co-authors. Institution: Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA. Major MEDLINE topics: Brain Neoplasms, Glutarates, Glioma. 45 references. The full text is freely available (open access, CC BY 4.0). Funded by: Mildred-Scheel Career Center (Deutsche Krebshilfe), NIH HHS, NIBIB NIH HHS.
Abstract (original language)
Background and Purpose In vivo MRS of 2‐hydroxyglutarate (2HG) may provide diagnostic and monitoring biomarkers in isocitrate dehydrogenase (IDH)‐mutated glioma. A previous meta‐analysis has shown good diagnostic accuracy of TE‐optimized PRESS for IDH‐mutated glioma, but most studies feature IDH‐wildtype glioma as a comparison. However, when considering newly identified brain lesions that may mimic glioma, full characterization of its diagnostic utility should also consider the accuracy of 2HG measurement in nontumor tissue. Therefore, we tested how well TE‐optimized 2HG levels distinguish between IDH‐mutated glioma and nontumor tissue—here, normal‐appearing brain. We further examined the impact of different spectral modeling strategies (baseline stiffness, macromolecule inclusion, and basis set composition).
Methods
Forty‐eight patients with diagnosed/suspected IDH‐mutated glioma were enrolled. 3T MRS data were acquired from tumor and contralateral nontumor tissue with TE‐optimized PRESS and analyzed with “LCModel” software. ROC analysis evaluated 2HG estimates’ ability to distinguish IDH‐mutated glioma from nontumor tissue. Modeling interactions between 2HG and other metabolites were evaluated to identify reasons for potential false‐positive 2HG detection.
Results
TE‐optimized PRESS distinguished IDH‐mutated glioma from nontumor tissue with lower sensitivity (range 0.76–0.62) and specificity (0.85–0.78) than the literature suggests for IDH‐mutated versus IDH‐wildtype glioma. Strong negative correlations between gamma‐aminobutyric acid (GABA) and 2HG persisted across all modeling strategies and may lead to false‐positive 2HG detection in nontumor tissue. We further present a cautionary example from a patient on a ketogenic diet, showing that ketone body acetone can interfere with 2HG detection.
Conclusions
Spectral overlap with GABA and acetone can lead to false‐positive 2HG detection in nontumor tissue.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
- Institute of Neuroradiology and Cooperative Brain Imaging Center - CoBIC, University Hospital, Frankfurt am Main, Germany.
- University Cancer Center Frankfurt (UCT), Frankfurt am Main, Germany.
- German Cancer Research Center (DKFZ), Frankfurt am Main, Germany.
- Frankfurt Cancer Institute (FCI), Frankfurt am Main, Germany.
- Department of Oncology, Sidney Kimmel Comprehensive Cancer Center, Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
Funding
- Mildred-Scheel Career Center (Deutsche Krebshilfe)
- NIH HHS
- NIBIB NIH HHS
- NIA NIH HHS
- National Institutes of Health
Cite
APA
Alcicek, S., Simicic, D., Blair, L., Saint-Germain, M., Zöllner, H. J., Davies-Jenkins, C. W., Holdhoff, M., Laterra, J., Bettegowda, C., Schreck, K. C., Lin, D. D., Barker, P. B., Kamson, D. O., & Oeltzschner, G. (2026). Pitfalls in 2HG Detection With TE‐Optimized MRS at 3T. Journal of Neuroimaging. https://doi.org/10.1111/jon.70146
Vancouver
Alcicek S, Simicic D, Blair L, Saint-Germain M, Zöllner HJ, Davies-Jenkins CW, et al. Pitfalls in 2HG Detection With TE‐Optimized MRS at 3T. Journal of Neuroimaging. 2026. doi:10.1111/jon.70146
BibTeX
@article{Alcicek2026,
author = {Alcicek, Seyma and Simicic, Dunja and Blair, Lindsay and Saint-Germain, Max and Zöllner, Helge J and Davies-Jenkins, Christopher W and Holdhoff, Matthias and Laterra, John and Bettegowda, Chetan and Schreck, Karisa C and Lin, Doris D and Barker, Peter B and Kamson, David O and Oeltzschner, Georg},
title = {Pitfalls in 2HG Detection With TE‐Optimized MRS at 3T},
journal = {Journal of Neuroimaging},
year = {2026},
doi = {10.1111/jon.70146}
}
EUVIMED