EUVIMEDObjevte EUVIMED
Crossref

Optimisation of EDS quantitative f‐ratio method via ZAF correction

Jiayi Chen, Yue Sun, Liujia Tian, Yiling Huang, Xuemei Song, Fan Peng, Wei Zheng, Yi Zeng

Journal of Microscopy · 2026-07 · 303 (1) · 79-87

DOI: 10.1111/jmi.70095 ↗

At a glance

Journal article in Journal of Microscopy, published online on 2026-04-16. Written by Jiayi Chen and 7 co-authors. Institution: School of Energy and Materials, Shanghai Polytechnic University, Shanghai, China. 18 references. Funded by: National Key Research and Development Program of China.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

This study investigates the f‐ratio quantitative method applicable to Cold Field Emission Scanning Electron Microscope/Energy Dispersive X‐Ray Spectroscopy (CFE‐SEM/EDS) systems. To address the issue of cumbersome calculations arising from the reliance on real‐time simulation in traditional f‐ratio methods, this study proposes an efficient strategy of ‘preset curve + ZAF correction’. This method employs pre‐constructed ‘f‐ratio vs. concentration’ models for individual elements derived from standard samples to substitute the traditional real‐time simulation of the unknown sample, and introduces ZAF factors to effectively compensate for matrix effects, thereby achieving quantitative analysis of unknown samples with matrix differences. An iterative approach was implemented to resolve the challenge of determining the absorption parameter (), which arises due to the unknown concentration of the unknown sample. F‐ratio models were established based on the P–In, As–Ga, S–Zn, Se–Cd, and S–Fe systems. Quantitative f‐ratio analysis was subsequently performed on the binary system As–In and Se–Zn, as well as the ternary system S–As–Fe. The results demonstrate that the ZAF correction f‐ratio method achieves satisfactory precision.

Subjects

Author keywords

EDSMonte–Carlo simulationZAF correctionquantitative X‐ray microanalysisthe f‐ratio method

Authors and institutions

  • School of Energy and Materials, Shanghai Polytechnic University, Shanghai, China.
  • State Key Laboratory of High Performance Ceramics, Shanghai Institute of Ceramics, Chinese Academy of Sciences, Shanghai, China.

Funding

  • National Key Research and Development Program of China

Cite

APA

Chen, J., Sun, Y., Tian, L., Huang, Y., Song, X., Peng, F., Zheng, W., & Zeng, Y. (2026). Optimisation of EDS quantitative f‐ratio method via ZAF correction. Journal of Microscopy, 303(1), 79-87. https://doi.org/10.1111/jmi.70095

Vancouver

Chen J, Sun Y, Tian L, Huang Y, Song X, Peng F, et al. Optimisation of EDS quantitative f‐ratio method via ZAF correction. Journal of Microscopy. 2026;303(1):79-87. doi:10.1111/jmi.70095

BibTeX
@article{Chen2026,
  author = {Chen, Jiayi and Sun, Yue and Tian, Liujia and Huang, Yiling and Song, Xuemei and Peng, Fan and Zheng, Wei and Zeng, Yi},
  title = {Optimisation of EDS quantitative f‐ratio method via ZAF correction},
  journal = {Journal of Microscopy},
  year = {2026},
  volume = {303},
  number = {1},
  pages = {79--87},
  doi = {10.1111/jmi.70095}
}