Fachartikel in der Zeitschrift Journal of Gastroenterology and Hepatology, 2023. Verfasst von Simone Vieira Toledo Guadagnin und 6 weiteren Autor:innen. Beteiligte Einrichtung: Clinics Hospital, Federal University of Goias, Goiânia, Goias, Brazil. Hauptthemen laut MEDLINE: Disinfection, Equipment Contamination. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: Fundação de Amparo à Pesquisa do Estado de Goiás, Coordenação de Aperfeiçoamento de Pessoal de Nível Superior.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Aim Forced‐air drying (FAD) cabinets are recommended for storage of reprocessed endoscopes, but financial constraints prevent their universal application. The study aimed to determine bacterial contamination in flexible gastroscopes (FG) channels after storage, in a cabinet with filtered air and UV lights, but without FAD.

Methods

Eight FG in clinical use in an endoscopy service of a large Brazilian hospital were sampled: immediately “Time zero” ( N = 50), 12 h “Time 1” ( N = 25), and 60 h “Time 2” ( N = 25) after reprocessing. Following a flush‐brush‐flush of channels, 40‐mL sterile water and 3 cm of the brush were collected. Each sample was divided, filtered onto two 0.22‐μm membranes, and incubated in media without or with disinfectant neutralizer. Automated method was used for identification and antibiotic resistance test of isolated bacteria.

Results

Bacterial contamination in times “1” and “2” was 5.9 and 16.1 times greater than that of “Time zero,” respectively. Number of positive cultures in media with and without neutralizer was similar at times “1” and “2,” while media with neutralizer produced more positive cultures at “Time zero.” Most bacteria isolated at “Time 2” were Gram‐negative rods (52.3%) and showed resistance to one or more antibiotics (65%).

Conclusion

Bacterial contamination was detected on reprocessed FG stored in non‐FAD cabinets overnight (12 h) and increased with longer storage time (60 h). The contamination source is likely to be bacteria in biofilm which multiply in the absence of FAD. Evidence‐based criteria should be available for storage time according to the cabinet available.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

DisinfectionEquipment ContaminationHumansEndoscopesBacteriaBrazil

Stichwörter der Autor:innen

DisinfectionEndoscopes, gastrointestinalEquipment contamination, biofilms

Autor:innen und Institutionen

  • Clinics Hospital, Federal University of Goias, Goiânia, Goias, Brazil.
  • Faculty of Nursing, Federal University of Goias, Goiânia, Goias, Brazil.
  • Faculty of Medicine, Health and Human Sciences, Macquarie University, Sydney, New South Wales, Australia.
  • Institute of Tropical Pathology and Public Health, Federal University of Goias, Goiânia, Goias, Brazil.

ORCID-Profile: Simone Vieira Toledo Guadagnin, Dayane de Melo Costa, Mariusa Gomes Borges Primo, Adriana Azevedo Silva, Lara Stefânia Netto de Oliveira Leão-Vasconcelos, Karen Vickery, Anaclara Ferreira Veiga Tipple

Förderung

  • Fundação de Amparo à Pesquisa do Estado de Goiás
  • Coordenação de Aperfeiçoamento de Pessoal de Nível Superior

Zitieren

APA

Guadagnin, S. V. T., Costa, D. M., Primo, M. G. B., Silva, A. A., Leão-Vasconcelos, L. S. N. O., Vickery, K., & Tipple, A. F. V. (2023). Significant increased bacterial contamination with endoscope overnight and weekend storage times. Journal of Gastroenterology and Hepatology. https://doi.org/10.1111/jgh.16224

Vancouver

Guadagnin SVT, Costa DM, Primo MGB, Silva AA, Leão-Vasconcelos LSNO, Vickery K, et al. Significant increased bacterial contamination with endoscope overnight and weekend storage times. Journal of Gastroenterology and Hepatology. 2023. doi:10.1111/jgh.16224

BibTeX
@article{Guadagnin2023,
  author = {Guadagnin, Simone Vieira Toledo and Costa, Dayane de Melo and Primo, Mariusa Gomes Borges and Silva, Adriana Azevedo and Leão-Vasconcelos, Lara Stefânia Netto de Oliveira and Vickery, Karen and Tipple, Anaclara Ferreira Veiga},
  title = {Significant increased bacterial contamination with endoscope overnight and weekend storage times},
  journal = {Journal of Gastroenterology and Hepatology},
  year = {2023},
  doi = {10.1111/jgh.16224}
}
Kostenlos starten →