Epicardial Pulsed Field Ablation Across the Intact Pericardium: Feasibility, Safety Margins, and Tissue Selectivity in a Porcine Model

Chin‐Yu Lin, Ting‐Yung Chang, Pei‐Yi Wang, Yung‐Nan Tsai, Shih‐Lin Chang, Ding‐Jun Juan, Cheng‐Han Chan, Yuen Hoong Phang, Arlene Melissa T. Dychiching, Shonda Ng

Journal of Cardiovascular Electrophysiology · 17.11.2026

DOI: 10.1111/jce.70565 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 17.11.2026. Verfasst von Chin‐Yu Lin und 9 weiteren Autor:innen. Die Arbeit nennt 19 Literaturangaben. Gefördert von: Biosense Webster, Taipei Veterans General Hospital.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Epicardial pulsed field ablation (PFA) holds promise, but in vivo safety, dose−responses, and the intact pericardium's barrier effect remain largely unexplored. Methods and Results Six pigs received 5, 10, 15, or 30 PFA shots via a pentaspline catheter (FARAWAVE system, 2.0 kV) placed externally on the intact left ventricular pericardium. Lesions were evaluated macroscopically and histologically at 1 month. PFA universally induced ventricular fibrillation requiring rescue shocks, and an observed numerical decline in left ventricular ejection fraction was noted at 1 month (72.6% to 50.1%). A distinct, nonlinear dose−response emerged. Standard doses (5–10 shots) yielded minimal true necrosis (0.15–2.63 cm 3 ) despite broader gross lesions. At 15 shots, true necrosis expanded to 4.39 cm 3 , approaching the safety threshold. Conversely, an extreme 30‐shot dose breached the pericardial impedance, causing a catastrophic spike in transmural necrosis (22.5 cm 3 ) and early lethal heart failure.

Conclusions

Epicardial PFA over an intact pericardium using a pentaspline catheter system could still create predictable lesions in the subepicardial area. The profound discrepancy between gross lesions and true necrosis suggests the pericardium acts as an electrical barrier. While massive overdoses cause non‐selective damage, PFA's ability to create scars through a biological membrane inspires future catheter designs that could trap electrolytic microbubbles to eliminate embolic risks.

Förderung

  • Biosense Webster
  • Taipei Veterans General Hospital

Zitieren

APA

Lin, C., Chang, T., Wang, P., Tsai, Y., Chang, S., Juan, D., Chan, C., Phang, Y. H., Dychiching, A. M. T., & Ng, S. (2026). Epicardial Pulsed Field Ablation Across the Intact Pericardium: Feasibility, Safety Margins, and Tissue Selectivity in a Porcine Model. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70565

Vancouver

Lin C, Chang T, Wang P, Tsai Y, Chang S, Juan D, et al. Epicardial Pulsed Field Ablation Across the Intact Pericardium: Feasibility, Safety Margins, and Tissue Selectivity in a Porcine Model. Journal of Cardiovascular Electrophysiology. 2026. doi:10.1111/jce.70565

BibTeX
@article{Lin2026,
  author = {Lin, Chin‐Yu and Chang, Ting‐Yung and Wang, Pei‐Yi and Tsai, Yung‐Nan and Chang, Shih‐Lin and Juan, Ding‐Jun and Chan, Cheng‐Han and Phang, Yuen Hoong and Dychiching, Arlene Melissa T. and Ng, Shonda},
  title = {Epicardial Pulsed Field Ablation Across the Intact Pericardium: Feasibility, Safety Margins, and Tissue Selectivity in a Porcine Model},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2026},
  doi = {10.1111/jce.70565}
}
Kostenlos starten →