Effect of Leadless Pacemaker Implantation on Tricuspid Regurgitation
Journal of Cardiovascular Electrophysiology · 2026-07 · 37 (7) · 1611-1614
DOI: 10.1111/jce.70386 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 01.06.2026. Verfasst von Scott D Eisenberg und 7 weiteren Autor:innen. Beteiligte Einrichtung: Emory University School of Medicine, Division of Internal Medicine, Atlanta, Georgia, USA. Hauptthemen laut MEDLINE: Tricuspid Valve Insufficiency, Pacemaker, Artificial, Cardiac Pacing, Artificial, Tricuspid Valve. Die Arbeit nennt 5 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Leadless pacemakers (LLPM) eliminate trans‐tricuspid leads and are presumed to reduce tricuspid regurgitation (TR), though longitudinal data are limited.
Methods
We retrospectively studied 356 patients with paired echocardiography before and after Micra implantation at a single tertiary care center. TR and MR severity were graded as none, mild, moderate, or severe. The primary analysis compared baseline with the most recent available echocardiogram. Secondary analyses examined TR severity at three mutually exclusive intervals: < 1 year, 1–3 years, and > 3 years postimplantation. Within‐patient changes were assessed using paired Wilcoxon signed‐rank tests.
Results
Over a median follow‐up of 540 days (IQR 67–1136 days), TR worsened in 155 patients (43.5%), improved in 89 (25.0%), and was unchanged in 112 (31.5%) at most recent available follow‐up ( p < 0.001). TR severity increased significantly at 1–3 years ( p = 0.003) and beyond 3 years ( p = 0.006), but not within the first year ( p = 0.12). The proportion with severe TR increased from 6.2% at baseline to 11.2% at most recent follow‐up, while none or mild TR declined from 75.5% to 65.2%. MR remained stable ( p = 0.14). Within‐patient comparisons showed greater TR than MR at all timepoints (all p < 0.001). Left ventricular ejection fraction declined modestly at all follow‐up intervals ( p < 0.001).
Conclusions
LLPM implantation was associated with progressive TR despite the absence of a trans‐tricuspid lead, with worsening emerging beyond the first year, supporting a remodeling‐based mechanism. The stability of MR argues against generalized valvular deterioration. Further studies are needed to define the clinical impact of LLPM‐associated TR progression.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Emory University School of Medicine, Division of Internal Medicine, Atlanta, Georgia, USA.
- Emory University School of Medicine, Division of Cardiovascular Disease, Atlanta, Georgia, USA.
- Division of Cardiac Electrophysiology, Emory University, Atlanta, Georgia, USA.
ORCID-Profile: Scott D Eisenberg, David R G Kulp, Mikhael F El-Chami
Zitieren
APA
Eisenberg, S. D., Evenhuis, B., Kulp, D. R. G., Weiss, A., Larson, J., Bhatia, N. K., El-Chami, M. F., & Merchant, F. M. (2026). Effect of Leadless Pacemaker Implantation on Tricuspid Regurgitation. Journal of Cardiovascular Electrophysiology, 37(7), 1611-1614. https://doi.org/10.1111/jce.70386
Vancouver
Eisenberg SD, Evenhuis B, Kulp DRG, Weiss A, Larson J, Bhatia NK, et al. Effect of Leadless Pacemaker Implantation on Tricuspid Regurgitation. Journal of Cardiovascular Electrophysiology. 2026;37(7):1611-1614. doi:10.1111/jce.70386
BibTeX
@article{Eisenberg2026,
author = {Eisenberg, Scott D and Evenhuis, Bernard and Kulp, David R G and Weiss, Allison and Larson, John and Bhatia, Neal K and El-Chami, Mikhael F and Merchant, Faisal M},
title = {Effect of Leadless Pacemaker Implantation on Tricuspid Regurgitation},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2026},
volume = {37},
number = {7},
pages = {1611--1614},
doi = {10.1111/jce.70386}
}
EUVIMED