Impact of Pulsed‐Field Ablation on the Left Atrial Appendage Diameter—Insights From Intra‐Procedural Echocardiography
Journal of Cardiovascular Electrophysiology · 2025
DOI: 10.1111/jce.70111 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2025. Verfasst von Nándor Szegedi und 13 weiteren Autor:innen. Beteiligte Einrichtung: Heart and Vascular Center, Semmelweis University, Budapest, Hungary. Hauptthemen laut MEDLINE: Atrial Appendage, Atrial Fibrillation, Catheter Ablation, Echocardiography, Transesophageal. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Project no. RRF-2.3.1-21-2022-00003 has been implemented with the support provided bythe European Union. Gábor Orbán and Nándor Szegedi are also supported by t….
Abstract (Originalsprache)
Introduction
Concomitant pulmonary vein isolation (PVI) and left atrial appendage (LAA) occlusion (LAAO) have become frequently used therapies. Pulsed‐field ablation (PFA), will likely be used for combined PVI plus LAAO procedures. However, there may be concerns regarding the malposition of the LAAO device attributed to the potential tissue edema after PVI.
Aim
We aimed to compare the LAA's size before and after PVI performed with the pentaspline catheter, measured intraprocedural by intracardiac echocardiography or transesophageal echocardiography.
Methods
We conducted a multicenter, prospective, observational study investigating PVI using the Farapulse system. The anteroposterior diameter of the left‐sided pulmonary vein (LPV), left atrial ridge, and LAA were measured before and after the PFA.
Results
We enrolled 91 patients aged 63 ± 10 years, 36% were women, and 47% had paroxysmal AF. The most common comorbidities was hypertension (59%). Procedure time and left atrial dwell time were 65 (52–80) min, and 27 (24–32) min, respectively. The diameter of the LAA was not different before and after the PVI (15 [13–18] and 16 [13–19], respectively; p = 0.756). On the other hand, the diameter of the left atrial ridge (7 [6–8] and 8 [6–8]) and the LPV (13 [10–15] and 13 [11–15]) was smaller before ablation compared to the diameter after PVI ( p < 0.0001 for both). No major complications occurred.
Conclusion
LAA anteroposterior diameter does not change after PVI with PFA. Although there is a significant change in the anteroposterior diameters of the LPV and the left atrial ridge, it does not seem clinically relevant. If significant edema formation is detected in a single case after PVI, postponing the LAAO procedure should be considered.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Heart and Vascular Center, Semmelweis University, Budapest, Hungary.
- Division of Clinical Electrophysiology, Department of Internal Medicine and Cardiology, Medical University of Warsaw, Warsaw, Poland.
- Liverpool Heart and Chest Hospital, Liverpool, UK.
ORCID-Profile: Nándor Szegedi, Piotr Gardziejczyk, Zoltán Salló, Patrik Tóth, Ferenc Komlósi, Ewa Wlazłowska-Struzik, Jakub Baran
Förderung
- Project no. RRF-2.3.1-21-2022-00003 has been implemented with the support provided bythe European Union. Gábor Orbán and Nándor Szegedi are also supported by t…
Zitieren
APA
Szegedi, N., Gardziejczyk, P., Salló, Z., Orbán, G., Boga, M., Szymkiewicz, P., Skowrońska, M., Tóth, P., Komlósi, F., Merkely, B., Wlazłowska-Struzik, E., Gellér, L., Baran, J., & Gupta, D. (2025). Impact of Pulsed‐Field Ablation on the Left Atrial Appendage Diameter—Insights From Intra‐Procedural Echocardiography. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.70111
Vancouver
Szegedi N, Gardziejczyk P, Salló Z, Orbán G, Boga M, Szymkiewicz P, et al. Impact of Pulsed‐Field Ablation on the Left Atrial Appendage Diameter—Insights From Intra‐Procedural Echocardiography. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.70111
BibTeX
@article{Szegedi2025,
author = {Szegedi, Nándor and Gardziejczyk, Piotr and Salló, Zoltán and Orbán, Gábor and Boga, Márton and Szymkiewicz, Paweł and Skowrońska, Marta and Tóth, Patrik and Komlósi, Ferenc and Merkely, Béla and Wlazłowska-Struzik, Ewa and Gellér, László and Baran, Jakub and Gupta, Dhiraj},
title = {Impact of Pulsed‐Field Ablation on the Left Atrial Appendage Diameter—Insights From Intra‐Procedural Echocardiography},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2025},
doi = {10.1111/jce.70111}
}
EUVIMED