Hemolysis Induced by Pulsed‐Field Ablation of Atrial Arrhythmias: A Comparative Analysis of Current Systems
Journal of Cardiovascular Electrophysiology · 2025-10 · 36 (10) · 2498-2506
DOI: 10.1111/jce.70049 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Cardiovascular Electrophysiology, online erschienen am 07.08.2025. Verfasst von Johannes Bruss und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Hauptthemen laut MEDLINE: Hemolysis, Catheter Ablation, Cardiac Catheters, Atrial Fibrillation. Die Arbeit wurde bisher 20-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: The authors received no specific funding for this work..
Abstract (Originalsprache)
Introduction
Pulsed‐field ablation (PFA) is an emerging technology associated with dose‐dependent hemolysis as a recently recognized side effect. This study aimed to compare hemolysis levels and assess dose‐dependency across three PFA systems: a pentaspline catheter (PSC), a lattice‐tip focal catheter (LTFC), and a variable loop circular catheter (VLCC).
Methods
Patients treated for atrial arrhythmias with the LTFC ( n = 29) or the VLCC ( n = 30) were included from a prospective registry. A matched cohort of patients treated with the PSC ( n = 28) was recruited from the same registry. Creatinine levels and markers of hemolysis were measured pre‐ablation and 1 day postablation.
Results
Haptoglobin levels decreased significantly more with the PSC and VLCC compared to the LTFC (−0.65 [−0.76, −0.49] g/L; −0.56 [−0.78, −0.43] g/L, −0.21 [−0.32, −0.1] g/L, respectively; p < 0.001 for both). Per‐application decreases in haptoglobin also differed (−17.5 [−20.38, −13.58] mg/L, −24.35 [−36.36, −17.92] mg/L, −3.61 [−5.98, −2.13] mg/L, respectively; p < 0.001 for both). There was no significant difference in haptoglobin decrease between the PSC and VLCC per procedure ( p = 1.0). Haptoglobin decrease per application was significantly larger with the VLCC compared to the PSC ( p = 0.0048). Per procedure LDH increase followed a similar trend (49 [18, 81.25] U/L; 14 [6, 60] U/L; 13 [−4, 46] U/L; respectively; p = 0.037). No hemolysis‐related complications were observed.
Conclusions
Hemolysis levels vary significantly among PFA platforms. Focal PFA catheters induce less hemolysis per procedure and application compared to large‐footprint catheters.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
- Sitem Center for Translational Medicine and Biomedical Entrepreneurship, University of Bern, Bern, Switzerland.
ORCID-Profile: Johannes Bruss, Thomas Kueffer, Andreas Haeberlin, Tobias Reichlin, Laurent Roten
Förderung
- The authors received no specific funding for this work.
Zitieren
APA
Bruss, J., Kueffer, T., Tanner, H., Noti, F., Haeberlin, A., Thalmann, G., Kozhuharov, N. A., Kovacs, B., Spahiu, V., Siklody, C. H., Reichlin, T., & Roten, L. (2025). Hemolysis Induced by Pulsed‐Field Ablation of Atrial Arrhythmias: A Comparative Analysis of Current Systems. Journal of Cardiovascular Electrophysiology, 36(10), 2498-2506. https://doi.org/10.1111/jce.70049
Vancouver
Bruss J, Kueffer T, Tanner H, Noti F, Haeberlin A, Thalmann G, et al. Hemolysis Induced by Pulsed‐Field Ablation of Atrial Arrhythmias: A Comparative Analysis of Current Systems. Journal of Cardiovascular Electrophysiology. 2025;36(10):2498-2506. doi:10.1111/jce.70049
BibTeX
@article{Bruss2025,
author = {Bruss, Johannes and Kueffer, Thomas and Tanner, Hildegard and Noti, Fabian and Haeberlin, Andreas and Thalmann, Gregor and Kozhuharov, Nikola Asenov and Kovacs, Boldizsar and Spahiu, Valon and Siklody, Claudia Herrera and Reichlin, Tobias and Roten, Laurent},
title = {Hemolysis Induced by Pulsed‐Field Ablation of Atrial Arrhythmias: A Comparative Analysis of Current Systems},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2025},
volume = {36},
number = {10},
pages = {2498--2506},
doi = {10.1111/jce.70049}
}
EUVIMED