Relation Between Instability of Ventricular Fibrillation Detection During Defibrillation Testing and Sudden Death in Future
Journal of Cardiovascular Electrophysiology · 2025
DOI: 10.1111/jce.16790 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2025. Verfasst von Sou Otsuki und 8 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Medical and Dental Sciences, Niigata University, Niigata, Japan. Hauptthemen laut MEDLINE: Ventricular Fibrillation, Defibrillators, Implantable, Death, Sudden, Cardiac, Electric Countershock. Die Arbeit nennt 23 Literaturangaben.
Abstract (Originalsprache)
Background
Sudden death (SD) due to ventricular tachycardia or fibrillation (VT/VF) detection failure in patients with implantable cardioverter‐defibrillators (ICDs) has been reported, despite its rarity. However, the relationship between VF detection instability during defibrillation testing (DT) and SD risk is unknown.
Objective
To investigate whether VF sensing status during DT predicts SD and VT/VF detection failure in patients with ICD/cardiac resynchronization therapy with defibrillator (CRT‐D).
Methods
Consecutive 97 patients with ICD/CRT‐D (Medtronic) who underwent DT were identified retrospectively. “Induced VF with detection failure” was defined as an episode with at least four ventricular sensed beats out of the VF detection interval in the most recent 18 sensed beats during DT. Patients were divided into two groups (detection‐failure or appropriate‐detection group) according to the presence of “induced VF with detection failure”. We investigated whether “induced VF with detection failure” predicts SD and inappropriate treatment interruption for clinically occurring VT/VF.
Results
Induced VF with detection failure was found in 8 of the 97 (8%) patients. During a median follow‐up of 9.9 years (interquartile range: 7.2–11.8 years), 9 SD occurred. Univariate analysis showed induced VF with detection failure was the only predictor for SD (hazard ratio, 7.94, 95% confidence interval, 1.96 − 32.1, p = 0.004). During the follow‐up, clinical VT/VF occurred in 41 patients. SD with inappropriate treatment interruption for a clinical VT/VF was observed in only three patients, two of them were in the detection‐failure group [detection‐failure: 2/5 (40%), appropriate‐detection: 1/36 (3%)].
Conclusion
The instability of VF detection in DT may predict the uncertainty of detection during clinically occurring VT/VF that causes SD.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Graduate School of Medical and Dental Sciences, Niigata University, Niigata, Japan.
- School of Medicine, Niigata University, Niigata, Japan.
ORCID-Profile: Sou Otsuki, Yuki Hasegawa, Masaomi Chinushi
Zitieren
APA
Otsuki, S., Sakurazawa, M., Furuse, H., Akagawa, R., Suzuki, N., Ikami, Y., Hasegawa, Y., Chinushi, M., & Inomata, T. (2025). Relation Between Instability of Ventricular Fibrillation Detection During Defibrillation Testing and Sudden Death in Future. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16790
Vancouver
Otsuki S, Sakurazawa M, Furuse H, Akagawa R, Suzuki N, Ikami Y, et al. Relation Between Instability of Ventricular Fibrillation Detection During Defibrillation Testing and Sudden Death in Future. Journal of Cardiovascular Electrophysiology. 2025. doi:10.1111/jce.16790
BibTeX
@article{Otsuki2025,
author = {Otsuki, Sou and Sakurazawa, Minori and Furuse, Hironori and Akagawa, Rie and Suzuki, Naomasa and Ikami, Yasuhiro and Hasegawa, Yuki and Chinushi, Masaomi and Inomata, Takayuki},
title = {Relation Between Instability of Ventricular Fibrillation Detection During Defibrillation Testing and Sudden Death in Future},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2025},
doi = {10.1111/jce.16790}
}
EUVIMED