Real‐world battery longevity of implantable loop recorders implanted for unexplained syncope: Results from a large single‐center registry
Journal of Cardiovascular Electrophysiology · 2024-11 · 35 (11) · 2134-2143
DOI: 10.1111/jce.16420 ↗At a glance
Beobachtungsstudie in Journal of Cardiovascular Electrophysiology, published online on 2024-09-09. Written by Lukas Duvillier and 5 co-authors. Institution: Department of Cardiology, Heart Center, Ghent University Hospital, Ghent, Belgium. Major MEDLINE topics: Syncope, Electrocardiography, Ambulatory, Predictive Value of Tests, Registries. Cited 5 times so far (Crossref); 44 references.
Abstract (original language)
Background/Purpose Implantable loop recorders (ILR) are increasingly used in cardiac rhythm monitoring and diagnostic work‐up of unexplained syncope. ILR battery longevity, according to manufacturers' product performance specifications, typically ranges between 2 and 4 years, but real‐world data in this population are lacking.
Methods
This monocentric, prospective, observational study included consecutive patients with unexplained syncope undergoing ILR implantation between October 2007 and 2019. The main purpose was to determine real‐world battery longevity of ILRs. Diagnostic yield and relationship between arrhythmogenic diagnosis and duration of ILR monitoring were explored.
Results
The study included 309 patients (59 years [38–73], 49% female) with ILR implantation for unexplained syncope. Median battery longevity was 42 [40–45] months. A total of 99.5% of ILRs reached prespecified battery longevity. The time to end‐of‐life varied by up to 33 months among the same ILR models. Overall arrhythmogenic diagnostic yield counted 27% (73% sick sinus syndrome, 20% atrioventricular block, and 7% ventricular tachycardia). Median time to diagnosis was 10 [2–25] months, with the latest event at 43 months. The cumulative diagnostic yield for arrhythmogenic event explaining syncope was 4.2%, 6.1%, 9.4%, 14.6%, 19.4%, and 26.7% at 1, 2, 6, 12, 24, and 48 months, respectively. In univariate analysis, first degree AV block and prolonged HV time on EP study were predictors of diagnosis, while QRS duration abnormality borderline missed significance.
Conclusions
Real‐world battery longevity of ILRs matched industry projected longevity in 99.5% of patients implanted with ILR for unexplained syncope. A battery longevity of minimum 3.5 years is recommended to maximize the diagnostic yield in this population.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Cardiology, Heart Center, Ghent University Hospital, Ghent, Belgium.
ORCID profiles: Lukas Duvillier, Jan De Pooter
Cite
APA
Duvillier, L., Demolder, A., Van Renterghem, S., De Mey, C., West, F., & De Pooter, J. (2024). Real‐world battery longevity of implantable loop recorders implanted for unexplained syncope: Results from a large single‐center registry. Journal of Cardiovascular Electrophysiology, 35(11), 2134-2143. https://doi.org/10.1111/jce.16420
Vancouver
Duvillier L, Demolder A, Van Renterghem S, De Mey C, West F, De Pooter J. Real‐world battery longevity of implantable loop recorders implanted for unexplained syncope: Results from a large single‐center registry. Journal of Cardiovascular Electrophysiology. 2024;35(11):2134-2143. doi:10.1111/jce.16420
BibTeX
@article{Duvillier2024,
author = {Duvillier, Lukas and Demolder, Anthony and Van Renterghem, Sally and De Mey, Caroline and West, Filip and De Pooter, Jan},
title = {Real‐world battery longevity of implantable loop recorders implanted for unexplained syncope: Results from a large single‐center registry},
journal = {Journal of Cardiovascular Electrophysiology},
year = {2024},
volume = {35},
number = {11},
pages = {2134--2143},
doi = {10.1111/jce.16420}
}