Durable pulmonary vein isolation with optimized high‐power and very high‐power short‐duration temperature‐controlled ablation: A step‐by‐step guide

Tom J R De Potter iD, Benjamin De Becker iD, Mattias Duytschaever iD

Journal of Cardiovascular Electrophysiology · 2024

DOI: 10.1111/jce.16217 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Cardiovascular Electrophysiology, 2024. Verfasst von Tom J R De Potter und 2 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Center, Division of Cardiology, Onze Lieve Vrouwziekenhuis Hospital, Aalst, Belgium. Hauptthemen laut MEDLINE: Atrial Fibrillation, Cardiac Catheters, Catheter Ablation, Equipment Design. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 41 Literaturangaben. Gefördert von: Biosense Webster, Inc..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Through systematic scientific rigor, the CLOSE guided workflow was developed and has been shown to improve pulmonary vein isolation durability. However, this technique was developed at a time when using power‐controlled ablation catheters with conventional power ranges was the norm. There has been increased adoption of a high‐power and very high‐power short‐duration ablation practice propelled by the availability of the temperature‐controlled radiofrequency QDOT MICRO catheter.

Methods

There are fundamental differences in biophysics between very high‐powered temperature guided ablation and conventional ablation strategy that may impact patient outcomes. The catheter's design and ablation modes offer flexibility in technique while accommodating the individual operator's clinical discretion and preference to deliver a durable, transmural, and contiguous lesion set.

Results

Here, we provide recommendations for 3 different workflows using the QDOT MICRO catheter in a step‐by‐step manner for pulmonary vein isolation based on our cumulative experience as early adopters of the technology and the data available in the scientific literature.

Conclusions

With standardization, temperature‐controlled ablation with the QDOT MICRO catheter provides operators the flexibility of implementing different ablation strategies to ensure durable contiguous pulmonary vein isolation depending on patient characteristics.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial FibrillationCardiac CathetersCatheter AblationEquipment DesignPulmonary VeinsHumansAction PotentialsHeart RateTemperatureTime FactorsTreatment OutcomeWorkflow

Stichwörter der Autor:innen

QDOT MICRO catheteratrial fibrillationcontact‐force sensingpulmonary vein isolationtemperature‐controlled ablationvery high‐power short duration

Autor:innen und Institutionen

  • Cardiovascular Center, Division of Cardiology, Onze Lieve Vrouwziekenhuis Hospital, Aalst, Belgium.
  • Department of Cardiology, AZ Sint-Jan Hospital, Ruddershove, Brugge, Belgium.

ORCID-Profile: Tom J R De Potter, Benjamin De Becker, Mattias Duytschaever

Förderung

  • Biosense Webster, Inc.

Zitieren

APA

De Potter, T. J. R., De Becker, B., & Duytschaever, M. (2024). Durable pulmonary vein isolation with optimized high‐power and very high‐power short‐duration temperature‐controlled ablation: A step‐by‐step guide. Journal of Cardiovascular Electrophysiology. https://doi.org/10.1111/jce.16217

Vancouver

De Potter TJR, De Becker B, Duytschaever M. Durable pulmonary vein isolation with optimized high‐power and very high‐power short‐duration temperature‐controlled ablation: A step‐by‐step guide. Journal of Cardiovascular Electrophysiology. 2024. doi:10.1111/jce.16217

BibTeX
@article{DePotter2024,
  author = {De Potter, Tom J R and De Becker, Benjamin and Duytschaever, Mattias},
  title = {Durable pulmonary vein isolation with optimized high‐power and very high‐power short‐duration temperature‐controlled ablation: A step‐by‐step guide},
  journal = {Journal of Cardiovascular Electrophysiology},
  year = {2024},
  doi = {10.1111/jce.16217}
}
Kostenlos starten →