EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A Novel Transcatheter Device for the Edge-to-Edge Treatment of Tricuspid Regurgitation: A Preliminary Evaluation

Eleonora Salurso, Francesca Perico, Fabio Pappalardo, Marco Gard, Matteo Antoniotti, Eugenio Passanante, Daniele Zanotti, Michele De Bonis, Ottavio Alfieri, Riccardo Vismara

Annals of Biomedical Engineering · 2024

DOI: 10.1007/s10439-023-03399-4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Annals of Biomedical Engineering, 2024. Verfasst von Eleonora Salurso und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electronics, Information and Bioengineering, Politecnico di Milano, Via Golgi 39, 20133, Milan, Italy. eleonora.salurso@polimi.it. Hauptthemen laut MEDLINE: Tricuspid Valve Insufficiency, Heart Valve Prosthesis Implantation. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Politecnico di Milano.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Tricuspid regurgitation (TR) is the most common pathology of the tricuspid valve (TV), with significant mortality in severe cases. A well-established strategy to treat TR is represented by the clover surgical technique, which consists of stitching together the free edges of TV leaflets, producing a clover-shaped valvular orifice. Transcatheter treatments for TR constitute a valuable alternative for high-risk patients. In this work we investigated haemodynamic performances and safety of a novel device (StarTric device (STD)) aiming to perform the clover technique via percutaneous access. To assess haemodynamic performances, STD and clover were applied on porcine pathological TVs and tested. Fluid dynamic indexes of both strategies were compared to the pathological model. To evaluate device safety, forces exchanged between device and leaflets were compared to the extraction force (EF) required to STD to completely pass through the leaflet. Clover technique and STD induced a comparable TV backflow reduction (48% and 47%, respectively), with associated increase of TV flow in all tested conditions. Diastolic transvalvular pressure similarly increased indicating a reduction, though not significant, of the valvular orifice. Forces ranged from 1N to 1.71N, compared to an EF of 22.16 ± 8.6N. Force varied significantly amongst different working conditions (normotensive, mild, and severe hypertensive) for each leaflet, whilst no significative variation was found on different leaflets in the same working condition. In the adopted experimental scenario, STD demonstrated comparable efficacy to the surgical strategy in restoring TV haemodynamic. The forces acting on the leaflets following STD implantation were far lower when compared to EFs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tricuspid Valve InsufficiencyHeart Valve Prosthesis ImplantationHumansAnimalsSwineTricuspid ValveHemodynamicsCathetersTreatment Outcome

Stichwörter der Autor:innen

Clover techniquePercutaneous treatmentsTricuspid regurgitation

Autor:innen und Institutionen

  • Department of Electronics, Information and Bioengineering, Politecnico di Milano, Via Golgi 39, 20133, Milan, Italy. eleonora.salurso@polimi.it.
  • Department of Electronics, Information and Bioengineering, Politecnico di Milano, Via Golgi 39, 20133, Milan, Italy.
  • Gard Consulting, Ivrea, Turin, Italy.
  • StarTric s.r.l., Milan, Italy.
  • Department of Cardiac Surgery, San Raffaele University Hospital, Milan, Italy.

Förderung

  • Politecnico di Milano

Zitieren

APA

Salurso, E., Perico, F., Pappalardo, F., Gard, M., Antoniotti, M., Passanante, E., Zanotti, D., De Bonis, M., Alfieri, O., & Vismara, R. (2024). A Novel Transcatheter Device for the Edge-to-Edge Treatment of Tricuspid Regurgitation: A Preliminary Evaluation. Annals of Biomedical Engineering. https://doi.org/10.1007/s10439-023-03399-4

Vancouver

Salurso E, Perico F, Pappalardo F, Gard M, Antoniotti M, Passanante E, et al. A Novel Transcatheter Device for the Edge-to-Edge Treatment of Tricuspid Regurgitation: A Preliminary Evaluation. Annals of Biomedical Engineering. 2024. doi:10.1007/s10439-023-03399-4

BibTeX
@article{Salurso2024,
  author = {Salurso, Eleonora and Perico, Francesca and Pappalardo, Fabio and Gard, Marco and Antoniotti, Matteo and Passanante, Eugenio and Zanotti, Daniele and De Bonis, Michele and Alfieri, Ottavio and Vismara, Riccardo},
  title = {A Novel Transcatheter Device for the Edge-to-Edge Treatment of Tricuspid Regurgitation: A Preliminary Evaluation},
  journal = {Annals of Biomedical Engineering},
  year = {2024},
  doi = {10.1007/s10439-023-03399-4}
}
Kostenlos starten →